Development and validation of a novel 7α-hydroxy-4-cholesten-3-one (C4) liquid chromatography tandem mass spectrometry method and its utility to assess pre-analytical stability.
Atkins, Jonathan S; Keevil, Brian G; Taylor, Angela E; Ludwig, Christian; Hawley, James M.
Afiliación
  • Atkins JS; Department of Clinical Biochemistry, Wythenshawe Hospital, Manchester University NHS Foundation Trust, Wythenshawe, UK.
  • Keevil BG; Department of Clinical Biochemistry, Wythenshawe Hospital, Manchester University NHS Foundation Trust, Wythenshawe, UK.
  • Taylor AE; Institute of Metabolism and Systems Research, The University of Birmingham, Birmingham, UK.
  • Ludwig C; Institute of Metabolism and Systems Research, The University of Birmingham, Birmingham, UK.
  • Hawley JM; Department of Clinical Biochemistry, Wythenshawe Hospital, Manchester University NHS Foundation Trust, Wythenshawe, UK.
Clin Chem Lab Med ; 2024 Aug 05.
Article en En | MEDLINE | ID: mdl-39097844
ABSTRACT

OBJECTIVES:

7α-Hydroxy-4-cholesten-3-one (C4) is the common intermediary of both primary bile acids. C4 is recommended by the British Society of Gastroenterology for the investigation of bile acid diarrhoea (BAD) in patients with chronic diarrhoea. This project aimed to develop and validate an assay to quantitate C4 in serum and assess the stability of C4 in unseparated blood.

METHODS:

Accuracy was underpinned by calibrating to quantitative nuclear magnetic resonance analysis. C4 was analysed in a 96-well plate format with a deuterated C4 internal standard and liquid-liquid extraction. Validation followed the 2018 Food and Drug Administration guidelines. To assess C4 stability, healthy volunteers (n=12) donated 8 fasted samples each. Samples were incubated at 20 °C for up to 72 h and retrieved, centrifuged, aliquoted and frozen for storage at different time points prior to C4 analysis.

RESULTS:

The C4 method demonstrated excellent analytical performance and passed all validation criteria. The method was found to be accurate, precise, free from matrix effects and interference. After 72 h of delayed sample separation, C4 concentration gradually declined by up to 14 % from baseline. However, the change was not significant for up to 12 h.

CONCLUSIONS:

We present a robust method of analysing serum C4, offering a convenient alternative to 75SeHCAT for BAD investigation. C4 was found to decline in unseparated blood over time; however, after 12 h the mean change was <5 % from baseline. Our results suggest C4 is suitable for collection from both primary and secondary care prior to gastroenterology referral.