Your browser doesn't support javascript.
loading
An improved signal processor for the ultrasonic dimension gauge.
Cardiovasc Res ; 14(8): 490-4, 1980 Aug.
Article en En | MEDLINE | ID: mdl-7438151
ABSTRACT
The flip-flop producing a pulse proportional to the transit time of an ultrasonic burst between two piezoelectric crystals was often reset by transmission artefacts in sonomicrometers. A design improvement was made. The insertion of two monostable vibrators and one nand gate corrected the resetting caused by transmission artefacts. The new design produced a flip-flop pulse that was proportional to the transit time of the ultrasonic burst through the aorta. Ascending aortic diameter was monitored in unanaesthetised dogs for 6 to 8 weeks using the new design.
Asunto(s)
Buscar en Google
Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Aorta / Ultrasonido / Ultrasonografía Tipo de estudio: Diagnostic_studies Límite: Animals Idioma: En Revista: Cardiovasc Res Año: 1980 Tipo del documento: Article
Buscar en Google