Your browser doesn't support javascript.
loading
The 62 MeV proton beam for the treatment of ocular melanoma at Clatterbridge.
Bonnett, D E; Kacperek, A; Sheen, M A; Goodall, R; Saxton, T E.
Afiliación
  • Bonnett DE; MRC Cyclotron Unit, Clatterbridge Hospital, Bebington, Merseyside, UK.
Br J Radiol ; 66(790): 907-14, 1993 Oct.
Article en En | MEDLINE | ID: mdl-8220975
ABSTRACT
A second treatment room and beam line has been constructed at the Cyclotron Unit at Clatterbridge for the purpose of using 62 MeV protons for the treatment of ocular melanoma. A uniform beam is produced by a double foil scattering system. The initial Bragg peak is spread across the target volume by the use of beam modulators. These are rotating four-vaned stepped absorbers made from Perspex. Two beam lines can be configured with different positions of modulators and range limiters. The first has a maximum penetration of 31.9 +/- 0.2 mm in water and the second a penetration of 31.2 +/- 0.2 mm. The second configuration has the advantage of less variation in beam penumbra, with a typical value of 1.7 +/- 0.1 mm for the 90% to 10% decrement lines. The patients are treated with individually shaped collimators. Beam output varies by less than 2% over the range of collimator areas used. The resulting whole-body dose equivalent to patient has also been assessed. In the first three years of operation over 250 patients have been treated.
Asunto(s)
Buscar en Google
Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Radioterapia de Alta Energía / Neoplasias del Ojo / Melanoma Límite: Humans Idioma: En Revista: Br J Radiol Año: 1993 Tipo del documento: Article País de afiliación: Reino Unido
Buscar en Google
Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Radioterapia de Alta Energía / Neoplasias del Ojo / Melanoma Límite: Humans Idioma: En Revista: Br J Radiol Año: 1993 Tipo del documento: Article País de afiliación: Reino Unido