ICD malfunction caused by excessive defibrillation charge time.
de Meester, A; Gilles, P H.
Afiliação
  • de Meester A; Cardiology Department, Jolimont Hospital, 7100 Haine Saint Paul, Belgium.
Pacing Clin Electrophysiol ; 24(4 Pt 1): 518-9, 2001 Apr.
Article em En | MEDLINE | ID: mdl-11341094
A 71-year-old man with a history of myocardial infarction and syncopal ventricular tachycardia received an ICD. Progressively, increased defibrillation charge times of the ICD were observed. During a fast ventricular tachycardia, the patient collapsed. The underlying malfunction could be documented by ICD telemetric interrogation. An excessive charge time was recorded causing subsequent delayed therapy. The ICD was considered potentially ineffective and required a generator exchange. This case report underlines the possible factors affecting the ICD charge times. Excessive charge time must be considered as a rare, or underestimated, cause of ICD malfunction, requiring generator replacement.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Taquicardia Ventricular / Desfibriladores Implantáveis / Análise de Falha de Equipamento / Eletrocardiografia / Infarto do Miocárdio Tipo de estudo: Diagnostic_studies / Etiology_studies Limite: Aged / Humans / Male Idioma: En Revista: Pacing Clin Electrophysiol Ano de publicação: 2001 Tipo de documento: Article País de afiliação: Bélgica
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Taquicardia Ventricular / Desfibriladores Implantáveis / Análise de Falha de Equipamento / Eletrocardiografia / Infarto do Miocárdio Tipo de estudo: Diagnostic_studies / Etiology_studies Limite: Aged / Humans / Male Idioma: En Revista: Pacing Clin Electrophysiol Ano de publicação: 2001 Tipo de documento: Article País de afiliação: Bélgica