ABSTRACT

BACKGROUND:

We encountered two out of 100 cases in which implanted biodegradable plates and screws had fractured within 1 month postoperatively.

MATERIAL:

Failure of the material was confirmed through clinical symptoms, radiographs or CT findings. In addition, four specimens obtained from these two cases were examined with regard to their ultrastructure using scanning electron microscopy.

RESULTS:

Several principal patterns of the fractured surface were found (1) gradual cracking, i.e. 'circular stair' and, (2) tortuous threads, i.e. a wavy line. It is conceivable that the material may not have been hit by major sudden forces but a disproportion between the thread configuration and the drilled hole may have led to screw loosening and torsion. Subsequently, the threads were deformed in a 'wavy' manner, finally leading to cracking and fracture of plates and screws. Fractures of plates and screws due to these instabilities are thought to be distinguishable from material resorption.

CONCLUSION:

In the application of biodegradable materials, more than two screws per single bone segment should be used as a principle of plate-fixation technique in order to avoid a stability-compromising situation, particularly in the stress-bearing areas of the maxillofacial region. Moreover, three-dimensional fixation using more than two plates is recommended in the facial skeleton e.g. zygomatic tripod. Intermaxillary fixation should also be considered to reinforce initial stability in stress-bearing areas.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Placas Ósseas / Parafusos Ósseos / Falha de Prótese / Implantes Absorvíveis Tipo de estudo: Observational_studies Limite: Adolescent / Adult / Aged / Child / Child, preschool / Female / Humans / Infant / Male / Middle aged Idioma: En Revista: J Craniomaxillofac Surg Assunto da revista: ODONTOLOGIA Ano de publicação: 2003 Tipo de documento: Article País de afiliação: Japão
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Placas Ósseas / Parafusos Ósseos / Falha de Prótese / Implantes Absorvíveis Tipo de estudo: Observational_studies Limite: Adolescent / Adult / Aged / Child / Child, preschool / Female / Humans / Infant / Male / Middle aged Idioma: En Revista: J Craniomaxillofac Surg Assunto da revista: ODONTOLOGIA Ano de publicação: 2003 Tipo de documento: Article País de afiliação: Japão