Improving the resolution of dynamic intensity modulated radiation therapy delivery by reducing the multileaf collimator sampling distance.
Greer, Peter B; Beckham, Wayne A; Ansbacher, William; Mann, Rita K.
Afiliação
  • Greer PB; B.C. Cancer Agency, Vancouver Island Center, Victoria, B.C., Canada. peter.greer@mater.health.nsw.gov.au
Med Phys ; 30(10): 2793-803, 2003 Oct.
Article em En | MEDLINE | ID: mdl-14596316
ABSTRACT
The conformality of a dose distribution delivered by a multileaf collimator (MLC) for intensity modulated radiation therapy (IMRT) is limited in the direction perpendicular to leaf motion by the finite leaf width. Two methods of improving the resolution of IMRT intensity maps in this direction were investigated. In the first, the desired fluence distribution is considered to be sampled by the MLC, with the sampling distance being the center-to-center distance between the MLC leaves. The sampling distance is reduced below the leaf width by combining separate irradiations with a couch shift between them. This has been applied to static field therapy [Galvin et al., Int. J. Radiat. Oncol., Biol., Phys. 35, 89-94 (1996)], and was proposed for IMRT by Bortfeld et al. [Med. Phys. 27, 2494-2502 (2000)]. In the second method, two MLC component fluences, with leaf width L = 2deltay and offset by deltay, are combined to reproduce desired intensity bins with deltay width. The effect of MLC leaf sampling distance on dose resolution was quantified for both 1.0 and 0.5 cm MLC leaf widths, utilizing a high resolution bar-pattern fluence, an annular shaped fluence, and an intensity step-edge. Improvement in resolution was found for the 1.0 cm leaf width at a sampling distance of 0.5 cm, with only a small benefit for further reduction. For the 0.5 cm leaf width, a sampling distance of 0.25 cm resulted in a dose resolution that was nearly independent of direction.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Radioterapia Conformacional Tipo de estudo: Risk_factors_studies Idioma: En Revista: Med Phys Ano de publicação: 2003 Tipo de documento: Article País de afiliação: Canadá
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Radioterapia Conformacional Tipo de estudo: Risk_factors_studies Idioma: En Revista: Med Phys Ano de publicação: 2003 Tipo de documento: Article País de afiliação: Canadá