High-frequency piezopolymer transducers with a copper-clad polyimide backing layer.
IEEE Trans Ultrason Ferroelectr Freq Control
; 53(7): 1376-80, 2006 Jul.
Article
em En
| MEDLINE
| ID: mdl-16889345
The effect of a copper-clad polyimide (CCP) backing layer on piezopolymer transducer performance is evaluated. High-frequency, spherically curved polyvinylidene fluoride (PVDF) transducers with and without a CCP backing layer are electrically and acoustically tested. The results showed very similar operating characteristics. B-mode in vivo images of a mouse embryo also showed no qualitative differences, indicating that the CCP backing layer does not affect transducer performance.
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Tipo de estudo:
Qualitative_research
Idioma:
En
Revista:
IEEE Trans Ultrason Ferroelectr Freq Control
Assunto da revista:
MEDICINA NUCLEAR
Ano de publicação:
2006
Tipo de documento:
Article