Your browser doesn't support javascript.
loading
Radiofrequency hepatic ablation with internally cooled electrodes and hybrid applicators with distant saline infusion using an in vivo porcine model.
Burdío, F; Navarro, A; Berjano, E J; Burdío, J M; Gonzalez, A; Güemes, A; Sousa, R; Rufas, M; Cruz, I; Castiella, T; Lozano, R; Lequerica, J L; Grande, L.
Afiliação
  • Burdío F; Department of Surgery, Hospital del Mar, Barcelona, Spain. fburdio@hotmail.com
Eur J Surg Oncol ; 34(7): 822-30, 2008 Jul.
Article em En | MEDLINE | ID: mdl-18042500
ABSTRACT

AIMS:

Radiofrequency ablation (RFA) of tumors by means of internally cooled (ICE) or multitined expandable electrodes combined with infusion of saline into the tissue may improve results. Our aim was to determine the efficacy of a previously optimized hybrid ICE system (ICE combined with infusion of saline into the tissue at a distance of 2mm) in comparison with a conventional ICE cluster electrode in porcine liver in vivo.

METHODS:

A total of 32 RFA were performed on a total of 10 farm pigs using two RFA systems Group A (n=16) Cluster electrode. Group B (n=16) Hybrid system (with continuous infusion of 100ml/h of 20% NaCl at 2mm distance from the electrode shaft by an independent isolated needle). Livers were removed for macroscopic and histological assessment after the procedure. Coagulation volume, coagulation diameters, coefficient of variability (CV) of coagulation volume, sphericity ratio (SR), deposited power (DP), deposited energy (DE), deposited energy per coagulation volume (DEV) and rise of animal temperature during the procedure were compared and correlated among groups. Additionally, linear regression analysis was modeled to study the relationship between deposited energy and either coagulation volume and rise of animal temperature during the procedure in both groups.

RESULTS:

Both coagulation volume and short diameter of coagulation were significantly greater (p<0.05) in group B compared to group A (22.7+/-11.0 cm(3) and 3.1+/-0.7 cm vs. 13.5+/-7.7 cm(3) and 2.5+/-0.5 cm, respectively). A similar CV and SR was observed among groups (57.1% and 1.4+/-0.3 vs. 48.6% and 1.3+/-0.2 for groups B and A, respectively). In group B, DE and DP were more than double group A, but DEV was nearly twice as high (9782 J/cm(3) vs. 5342 J/cm(3), for groups B and A, respectively). No significant relationship between DE and coagulation volume was encountered.

CONCLUSION:

Efficacy of a single ICE may be improved with continuous infusion of saline at around 2 mm from the electrode shaft. Coagulation volume obtained with this improved system may be even greater than that obtained with a cluster electrode.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Ablação por Cateter / Neoplasias Hepáticas Tipo de estudo: Evaluation_studies Limite: Animals Idioma: En Revista: Eur J Surg Oncol Assunto da revista: NEOPLASIAS Ano de publicação: 2008 Tipo de documento: Article País de afiliação: Espanha
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Ablação por Cateter / Neoplasias Hepáticas Tipo de estudo: Evaluation_studies Limite: Animals Idioma: En Revista: Eur J Surg Oncol Assunto da revista: NEOPLASIAS Ano de publicação: 2008 Tipo de documento: Article País de afiliação: Espanha