Immunoglobulin G paraprotein cross-reactivity with immunoglobulin M measurement.
Hawkins, Robert C; Lee, Lian-King.
Afiliação
  • Hawkins RC; Department of Pathology and Laboratory Medicine, Tan Tock Seng Hospital, Singapore 308433. Robert_Hawkins@ttsh.com.sg
Ann Clin Biochem ; 45(Pt 4): 431-3, 2008 Jul.
Article em En | MEDLINE | ID: mdl-18583633
ABSTRACT
We report a case of IgG paraprotein cross-reactivity with IgM measurement in a patient with IgGkappa multiple myeloma and associated IgA and IgM immune paresis. IgM measurements using a Beckman-Coulter Synchron LX IgM reagent initially gave unmeasurably low IgM concentrations (<0.3 g/L) but IgM quantitation using later batches of reagent gave IgM concentrations of 13-42 g/L. Immunofixation confirmed the continued presence of the IgGkappa paraprotein band with suppression of IgA and IgM and analysis on the Behring BN2 and Beckman-Coulter Immage instruments showed IgM concentrations of <0.2 g/L. Immunofixation using the Synchron LX IgM reagent as antiserum confirmed binding of the Synchron LX IgM antiserum to the IgGkappa paraprotein. This case highlights the importance of checking all unexpected immunoglobulin measurements with previous results and immunofixation findings to avoid mistakes.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Imunoglobulina G / Imunoglobulina M / Paraproteínas Limite: Aged / Humans Idioma: En Revista: Ann Clin Biochem Ano de publicação: 2008 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Imunoglobulina G / Imunoglobulina M / Paraproteínas Limite: Aged / Humans Idioma: En Revista: Ann Clin Biochem Ano de publicação: 2008 Tipo de documento: Article