Enantioselective determination of thyroxine enantiomers by ligand-exchange CE with UV absorbance and ICP-MS detection.
Electrophoresis
; 30(10): 1774-82, 2009 May.
Article
em En
| MEDLINE
| ID: mdl-19441034
A simple CE method has been developed for the separation and determination of thyroxine (T4) enantiomers in pharmaceutical formulations. The method was based on ligand-exchange mechanism using a Cu(II)/L-proline complex as chiral selector. The effects of different parameters affecting separation such as chiral selector concentration, organic additive, buffer pH and temperature were investigated. A baseline separation of the two enantiomers was obtained at a Cu(II)/L-proline ratio of 1:8 in a borate buffer (15 mmol/L, pH 9.6) containing 10% v/v acetonitrile. Under the optimized conditions, precision linearity range and detection limits of the developed enantioselective CE method were evaluated and compared using two different detection systems: conventional UV detection at 226 nm and iodine (127I)specific detection ("chiral speciation") with ICP-MS. Both methodologies show adequate analytical performance characteristics with detection limits around 0.30 microg/mL for each enantiomer of T4. Finally, a levothroid pharmaceutical formulation sample was successfully analyzed using both developed methods CE-UV and CE-ICP-MS.
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Espectrofotometria Ultravioleta
/
Espectrometria de Massas
/
Tiroxina
/
Eletroforese Capilar
Tipo de estudo:
Diagnostic_studies
Idioma:
En
Revista:
Electrophoresis
Ano de publicação:
2009
Tipo de documento:
Article
País de afiliação:
Espanha