The System Design, Engineering Architecture, and Preliminary Results of a Lower-Cost High-Sensitivity High-Resolution Positron Emission Mammography Camera.
Zhang, Yuxuan; Ramirez, Rocio A; Li, Hongdi; Liu, Shitao; An, Shaohui; Wang, Chao; Baghaei, Hossain; Wong, Wai-Hoi.
Afiliação
  • Zhang Y; M. D. Anderson Cancer Center, University of Texas, Houston, TX 77030 USA.
IEEE Trans Nucl Sci ; 57(1): 104-110, 2010 Feb 01.
Article em En | MEDLINE | ID: mdl-20485539
ABSTRACT
A lower-cost high-sensitivity high-resolution positron emission mammography (PEM) camera is developed. It consists of two detector modules with the planar detector bank of 20 × 12 cm(2). Each bank has 60 low-cost PMT-Quadrant-Sharing (PQS) LYSO blocks arranged in a 10 × 6 array with two types of geometries. One is the symmetric 19.36 × 19.36 mm(2) block made of 1.5 × 1.5 × 10 mm(3) crystals in a 12 × 12 array. The other is the 19.36 × 26.05 mm(2) asymmetric block made of 1.5 × 1.9 × 10 mm(3) crystals in 12 × 13 array. One row (10) of the elongated blocks are used along one side of the bank to reclaim the half empty PMT photocathode in the regular PQS design to reduce the dead area at the edge of the module. The bank has a high overall crystal packing fraction of 88%, which results in a very high sensitivity. Mechanical design and electronics have been developed for low-cost, compactness, and stability purposes. Each module has four Anger-HYPER decoding electronics that can handle a count-rate of 3 Mcps for single events. A simple two-module coincidence board with a hardware delay window for random coincidences has been developed with an adjustable window of 6 to 15 ns. Some of the performance parameters have been studied by preliminary tests and Monte Carlo simulations, including the crystal decoding map and the 17% energy resolution of the detectors, the point source sensitivity of 11.5% with 50 mm bank-to-bank distance, the 1.2 mm-spatial resolutions, 42 kcps peak Noise Equivalent Count Rate at 7.0-mCi total activity in human body, and the resolution phantom images. Those results show that the design goal of building a lower-cost, high-sensitivity, high-resolution PEM detector is achieved.

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Tipo de estudo: Diagnostic_studies / Health_economic_evaluation Idioma: En Revista: IEEE Trans Nucl Sci Ano de publicação: 2010 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Tipo de estudo: Diagnostic_studies / Health_economic_evaluation Idioma: En Revista: IEEE Trans Nucl Sci Ano de publicação: 2010 Tipo de documento: Article