Comparison of gamma-ray coincidence and low-background gamma-ray singles spectrometry.
Appl Radiat Isot
; 70(2): 392-6, 2012 Feb.
Article
em En
| MEDLINE
| ID: mdl-22037206
ABSTRACT
Aerosol samples have been studied under different background conditions using gamma-ray coincidence and low-background gamma-ray singles spectrometric techniques with High-Purity Germanium detectors. Conventional low-background gamma-ray singles counting is a competitive technique when compared to the gamma-gamma coincidence approach in elevated background conditions. However, measurement of gamma-gamma coincidences can clearly make the identification of different nuclides more reliable and efficient than using singles spectrometry alone. The optimum solution would be a low-background counting station capable of both singles and gamma-gamma coincidence spectrometry.