Uncertainty of ozone measurements with the primary standard reference photometer (SRP45).
Talanta
; 86: 71-81, 2011 Oct 30.
Article
em En
| MEDLINE
| ID: mdl-22063513
ABSTRACT
A comparison of the ozone primary reference standard photometer serial number 45 (SRP45) against the National Institute of Standards and Technology (NIST) instruments, serial number 0 (SRP0) and 2 (SRP2), has been performed in order to establish the traceability and comparability of ozone measurements made by the Chilean atmospheric science community. A complete uncertainty budget was developed for SRP45, using a GUM approach. The results of the comparisons allow us to conclude that SRP45, SRP0 and SRP2 are comparable according to international criteria over an ozone mole fraction range of 0 nmol mol(-1) to at least 500 nmol mol(-1). The official result for the validation of SRP45 is x(ozone)(SRP45)=[0.013+0.99806x(ozone)(SRP2)] nmol mol(-1) with an expanded uncertainty of [Formula in text] from 0 to 500 nmol mol(-1).
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Idioma:
En
Revista:
Talanta
Ano de publicação:
2011
Tipo de documento:
Article
País de afiliação:
Chile