ABSTRACT
A comparison of the ozone primary reference standard photometer serial number 45 (SRP45) against the National Institute of Standards and Technology (NIST) instruments, serial number 0 (SRP0) and 2 (SRP2), has been performed in order to establish the traceability and comparability of ozone measurements made by the Chilean atmospheric science community. A complete uncertainty budget was developed for SRP45, using a GUM approach. The results of the comparisons allow us to conclude that SRP45, SRP0 and SRP2 are comparable according to international criteria over an ozone mole fraction range of 0 nmol mol(-1) to at least 500 nmol mol(-1). The official result for the validation of SRP45 is x(ozone)(SRP45)=[0.013+0.99806x(ozone)(SRP2)] nmol mol(-1) with an expanded uncertainty of [Formula in text] from 0 to 500 nmol mol(-1).

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Revista: Talanta Ano de publicação: 2011 Tipo de documento: Article País de afiliação: Chile