Your browser doesn't support javascript.
loading
Calibration of stack monitors for measurement of noble gases in nuclear facilities.
Kovar, Petr; Dryak, Pavel; Suran, Jiri; Gudelis, Arunas.
Afiliação
  • Kovar P; Czech Metrology Institute, Inspectorate for Ionizing Radiation, Radiova 1, CZ-102 00 Prague 10, Czech Republic. pkovar@cmi.cz
Appl Radiat Isot ; 70(9): 2127-9, 2012 Sep.
Article em En | MEDLINE | ID: mdl-22424839
ABSTRACT
In nuclear facilities stack monitors are used for the measurement of the volumetric activity of noble gases. Spectrometric measurement is needed because the content of stack effluents is always a mixture of radionuclides. In some nuclear power plants new types of monitors were installed based on HPGe detectors. For efficiency calibration a standard with the radionuclide Xe-127 was developed and calibration curve constructed in the energy range 81 keV-1293 keV. Experiental efficiencies were checked using an MC model.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Radiometria / Gases Nobres / Poluentes Radioativos do Ar / Centrais Nucleares Tipo de estudo: Prognostic_studies Idioma: En Revista: Appl Radiat Isot Assunto da revista: MEDICINA NUCLEAR / SAUDE AMBIENTAL Ano de publicação: 2012 Tipo de documento: Article País de afiliação: República Tcheca

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Radiometria / Gases Nobres / Poluentes Radioativos do Ar / Centrais Nucleares Tipo de estudo: Prognostic_studies Idioma: En Revista: Appl Radiat Isot Assunto da revista: MEDICINA NUCLEAR / SAUDE AMBIENTAL Ano de publicação: 2012 Tipo de documento: Article País de afiliação: República Tcheca