Sensitivity and accuracy of volumetry of pulmonary nodules on low-dose 16- and 64-row multi-detector CT: an anthropomorphic phantom study.
Xie, Xueqian; Zhao, Yingru; Snijder, Roland A; van Ooijen, Peter M A; de Jong, Pim A; Oudkerk, Matthijs; de Bock, Geertruida H; Vliegenthart, Rozemarijn; Greuter, Marcel J W.
Afiliação
  • Xie X; Department of Radiology, University of Groningen, University Medical Center Groningen, EB44, P.O. Box 30.001, 9700 RB, Groningen, The Netherlands.
Eur Radiol ; 23(1): 139-47, 2013 Jan.
Article em En | MEDLINE | ID: mdl-22797957
ABSTRACT

OBJECTIVE:

To assess the sensitivity of detection and accuracy of volumetry by manual and semi-automated quantification of artificial pulmonary nodules in an anthropomorphic thoracic phantom on low-dose CT.

METHODS:

Fifteen artificial spherical nodules (diameter 3, 5, 8, 10 and 12 mm; CT densities -800, -630 and +100 HU) were randomly placed inside an anthropomorphic thoracic phantom. The phantom was examined on 16- and 64-row multidetector CT with a low-dose protocol. Two independent blinded observers screened for pulmonary nodules. Nodule diameter was measured manually, and volume calculated. For solid nodules (+100 HU), diameter and volume were also evaluated by semi-automated software. Differences in observed volumes between the manual and semi-automated method were evaluated by a t-test.

RESULTS:

Sensitivity was 100 % for all nodules of >5 mm and larger, 60-80 % for solid and 0-20 % for non-solid 3-mm nodules. No false-positive nodules but high inter-observer reliability and inter-technique correlation were found. Volume was underestimated manually by 24.1 ± 14.0 % for nodules of any density, and 26.4 ± 15.5 % for solid nodules, compared with 7.6 ± 8.5 % (P < 0.01) semi-automatically.

CONCLUSION:

In an anthropomorphic phantom study, the sensitivity of detection is 100 % for nodules of >5 mm in diameter. Semi-automated volumetry yielded more accurate nodule volumes than manual measurements.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Tomografia Computadorizada por Raios X / Nódulo Pulmonar Solitário / Imagens de Fantasmas Tipo de estudo: Diagnostic_studies Limite: Humans Idioma: En Revista: Eur Radiol Assunto da revista: RADIOLOGIA Ano de publicação: 2013 Tipo de documento: Article País de afiliação: Holanda

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Tomografia Computadorizada por Raios X / Nódulo Pulmonar Solitário / Imagens de Fantasmas Tipo de estudo: Diagnostic_studies Limite: Humans Idioma: En Revista: Eur Radiol Assunto da revista: RADIOLOGIA Ano de publicação: 2013 Tipo de documento: Article País de afiliação: Holanda