Your browser doesn't support javascript.
loading
Reproducibility of optical coherence tomography measurements in children.
Altemir, Irene; Pueyo, Victoria; Elía, Noemi; Polo, Vicente; Larrosa, Jose M; Oros, Daniel.
Afiliação
  • Altemir I; Department of Ophthalmology, Hospital Universitario Miguel Servet, Zaragoza, Spain.
Am J Ophthalmol ; 155(1): 171-176.e1, 2013 Jan.
Article em En | MEDLINE | ID: mdl-22967864
ABSTRACT

PURPOSE:

To determine the interobserver and intraobserver reproducibility of a Fourier-domain optical coherence tomography device (Cirrus HD OCT; Carl Zeiss Meditec, Dublin, California, USA) in normal pediatric eyes.

DESIGN:

Prospective cross-sectional study.

METHODS:

One hundred healthy children were recruited prospectively and consecutively. Only 1 randomly chosen eye per subject was included in the study. The eye underwent 3 scans centered on the optic disc and another 3 scans centered on the macula that were acquired by a single operator. A fourth examination was performed by a second operator. Interobserver and intraobserver reproducibility were described by intraclass correlation coefficients (ICCs) and coefficients of variation (COVs).

RESULTS:

The mean age was 9.15 years (range, 6.22 to 11.31 years; standard deviation, 1.05 years). Mean retinal nerve fiber layer thickness was 99.53 µm (standard deviation, 10.10 µm), and mean macular thickness was 282.91 µm (standard deviation, 11.83 µm). All the parameters evaluated were highly reproducible. Intraobserver COVs of the retinal nerve fiber layer measurements ranged from 2.24% to 5.52%, and the COV of macular thickness was 0.97%. The intraclass correlation coefficient was greater than 0.8 for all the parameters. The interobserver COV ranged from 2.23% to 5.18%, and the COV of macular thickness was 0.82%. In all the evaluated parameters, the intraclass correlation coefficient was more than 0.75. Repeatability was slightly better in children older than 10 years than in children younger than 9 years.

CONCLUSIONS:

Retinal nerve fiber layer and macular measurements obtained by Fourier-domain optical coherence tomography showed good repeatability for healthy eyes in the pediatric population. Cirrus HD OCT examinations of the retina are reliable in children.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Disco Óptico / Células Ganglionares da Retina / Tomografia de Coerência Óptica / Técnicas de Diagnóstico Oftalmológico / Macula Lutea / Fibras Nervosas Tipo de estudo: Clinical_trials / Diagnostic_studies / Observational_studies / Prevalence_studies / Risk_factors_studies Limite: Child / Female / Humans / Male Idioma: En Revista: Am J Ophthalmol Ano de publicação: 2013 Tipo de documento: Article País de afiliação: Espanha

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Disco Óptico / Células Ganglionares da Retina / Tomografia de Coerência Óptica / Técnicas de Diagnóstico Oftalmológico / Macula Lutea / Fibras Nervosas Tipo de estudo: Clinical_trials / Diagnostic_studies / Observational_studies / Prevalence_studies / Risk_factors_studies Limite: Child / Female / Humans / Male Idioma: En Revista: Am J Ophthalmol Ano de publicação: 2013 Tipo de documento: Article País de afiliação: Espanha