Variations in daily quality assurance dosimetry from device levelling, feet position and backscatter material.
Ceylan, Abdurrahman; Butson, Martin; Cullen, Ashley; Yu, Peter K N; Alnawaf, Hani.
Afiliação
  • Ceylan A; Illawarra Shoalhaven Cancer Network, Wollongong Hospital, Wollongong, Australia.
Australas Phys Eng Sci Med ; 35(4): 485-9, 2012 Dec.
Article em En | MEDLINE | ID: mdl-23208619
ABSTRACT
Daily quality assurance procedures are an essential part of radiotherapy medical physics. Devices such as the Sun Nuclear, DQA3 are effective tools for analysis of daily dosimetry including flatness, symmetry, energy, field size and central axis radiation dose measurement. The DQA3 can be used on the treatment couch of the linear accelerator or on a dedicated table/bed for superficial and orthovoltage x-ray machines. This device is levelled using its dedicated feet. This work has shown that depending on the quantity of backscatter material behind the DQA3 device, the position of the levelling feet can affect the measured central axis dose by up to 1.8 % (250 kVp and 6 MV) and that the introduction of more backscatter material behind the DQA3 can lead to up to 7.2 % (6 MV) variations in measured central axis dose. In conditions where no backscatter material is present, dose measurements can vary up to 1 %. As such this work has highlighted the need to keep the material behind the DQA3 device constant as well as maintaining the accuracy of the feet position on the device to effectively measure the most accurate daily constancy achievable. Results have also shown that variations in symmetry and energy calculations of up to 1 % can occur if the device is not levelled appropriately. As such, we recommend the position of the levelling feet on the device be as close as possible to the device so that a constant distance is kept between the DQA3 and the treatment couch and thus minimal levelling variations also occur. We would also recommend having no extra backscattering material behind the DQA3 device during use to minimise any variations which might occur from these backscattering effects.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Garantia da Qualidade dos Cuidados de Saúde / Radiometria / Artefatos Tipo de estudo: Diagnostic_studies País/Região como assunto: Oceania Idioma: En Revista: Australas Phys Eng Sci Med Ano de publicação: 2012 Tipo de documento: Article País de afiliação: Austrália

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Garantia da Qualidade dos Cuidados de Saúde / Radiometria / Artefatos Tipo de estudo: Diagnostic_studies País/Região como assunto: Oceania Idioma: En Revista: Australas Phys Eng Sci Med Ano de publicação: 2012 Tipo de documento: Article País de afiliação: Austrália