Your browser doesn't support javascript.
loading
A high-throughput LC-MS/MS method suitable for population biomonitoring measures five serum folate vitamers and one oxidation product.
Fazili, Zia; Whitehead, Ralph D; Paladugula, Neelima; Pfeiffer, Christine M.
Afiliação
  • Fazili Z; Division of Laboratory Sciences, National Center for Environmental Health, Centers for Disease Control and Prevention, Atlanta, GA 30341, USA.
Anal Bioanal Chem ; 405(13): 4549-60, 2013 May.
Article em En | MEDLINE | ID: mdl-23462981
ABSTRACT
Small specimen volume and high sample throughput are key features needed for routine methods used for population biomonitoring. We modified our routine eight-probe solid phase extraction (SPE) LC-MS/MS method for the measurement of five folate vitamers [5-methyltetrahydrofolate (5-methylTHF), folic acid (FA), plus three minor forms THF, 5-formylTHF, 5,10-methenylTHF] and one oxidation product of 5-methylTHF (MeFox) to require less serum volume (150 µL instead of 275 µL) by using 96-well SPE plates with 50 mg instead of 100 mg phenyl sorbent and to provide faster throughput by using a 96-probe SPE system. Total imprecision (10 days, two replicates/day) for three serum quality control pools was 2.8-3.6% for 5-methylTHF (19.5-51.1 nmol/L), 6.6-8.7% for FA (0.72-11.4 nmol/L), and ≤11.4% for the minor folate forms (<1-5 nmol/L). The mean (±SE) recoveries of folates spiked into serum (3 days, four levels, two replicates/level) were 5-methylTHF, 99.4 ± 3.6%; FA, 100 ± 1.8%; minor folates, 91.7-108%. SPE extraction efficiencies were ≥85%, except for THF (78%). Limits of detection were ≤0.3 nmol/L. The new method correlated well with our routine method [n = 150, r = 0.99 for 5-methylTHF, FA, and total folate (tFOL, sum of folate forms)] and produced slightly higher tFOL (5.6%) and 5-methylTHF (7.3%) concentrations, likely due to the faster 96-probe SPE process (1 vs. 5 h), resulting in improved SPE efficiency and recovery compared to the eight-probe SPE method. With this improved LC-MS/MS method, 96 samples can be processed in ~2 h, and all relevant folate forms can be accurately measured using a small serum volume.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Tetra-Hidrofolatos / Leucovorina / Ensaios de Triagem em Larga Escala / Ácido Fólico Limite: Humans Idioma: En Revista: Anal Bioanal Chem Ano de publicação: 2013 Tipo de documento: Article País de afiliação: Estados Unidos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Tetra-Hidrofolatos / Leucovorina / Ensaios de Triagem em Larga Escala / Ácido Fólico Limite: Humans Idioma: En Revista: Anal Bioanal Chem Ano de publicação: 2013 Tipo de documento: Article País de afiliação: Estados Unidos