Your browser doesn't support javascript.
loading
Optimal composition of dialysis fluid for on-line hemodiafiltration in central dialysis fluid delivery system.
Tomo, Tadashi.
Afiliação
  • Tomo T; Department of Nephrology, Oita University Hospital, Oita, Japan. tomo@oita-u.ac.jp
Blood Purif ; 35 Suppl 1: 69-73, 2013.
Article em En | MEDLINE | ID: mdl-23466383
ABSTRACT
In this paper, we propose 'the preferred composition of dialysis fluid for on-line hemodiafiltration (HDF) in central dialysis fluid delivery system, (CDDS)' sodium concentration 138-140 mEq/l; potassium concentration 2.0 mEq/l; calcium concentration 2.5-3.5 mEq/l; glucose concentration 1.0-1.5 g/l; buffer bicarbonate (acetate free is preferred). Concentrations of electrolytes in dialysis fluid (= substitution fluid) are applied to the compositions of launched dialysis fluid in Japan. From the viewpoint of biocompatibility and correction of metabolic acidosis, acetate-free bicarbonate buffer is thought to be important. This proposal of the preferred composition of dialysis fluid for on-line HDF in CDDS is based on a few reports. Further investigations and research to reveal the optimal composition of dialysis fluid for on-line HDF are expected in the future.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Soluções para Hemodiálise / Hemodiafiltração Tipo de estudo: Etiology_studies Limite: Humans Idioma: En Revista: Blood Purif Ano de publicação: 2013 Tipo de documento: Article País de afiliação: Japão

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Soluções para Hemodiálise / Hemodiafiltração Tipo de estudo: Etiology_studies Limite: Humans Idioma: En Revista: Blood Purif Ano de publicação: 2013 Tipo de documento: Article País de afiliação: Japão