Your browser doesn't support javascript.
loading
Non invasive evaluation of cardiomechanics in patients undergoing MitrClip procedure.
Guarracino, Fabio; Ferro, Baldassare; Baldassarri, Rubia; Bertini, Pietro; Forfori, Francesco; Giannini, Cristina; Di Bello, Vitantonio; Petronio, Anna S.
Afiliação
  • Guarracino F; Department of Cardiothoracic Anesthesia and Intensive Care Medicine, Azienda Ospedaliero Universitaria Pisana, Via Paradisa 2, 56123, Italy. fabiodoc64@hotmail.com
Cardiovasc Ultrasound ; 11: 13, 2013 May 04.
Article em En | MEDLINE | ID: mdl-23642140
ABSTRACT

BACKGROUND:

In the last recent years a new percutaneous procedure, the MitraClip, has been validated for the treatment of mitral regurgitation. MitraClip procedure is a promising alternative for patients unsuitable for surgery as it reduces the risk of death related to surgery ensuring a similar result. Few data are present in literature about the variation of hemodynamic parameters and ventricular coupling after Mitraclip implantation.

METHODS:

Hemodynamic data of 18 patients enrolled for MitraClip procedure were retrospectively reviewed and analyzed. Echocardiographic measurements were obtained the day before the procedure (T0) and 21 ± 3 days after the procedure (T1), including evaluation of Ejection Fraction, mitral valve regurgitation severity and mechanism, forward Stroke Volume, left atrial volume, estimated systolic pulmonary pressure, non invasive echocardiographic estimation of single beat ventricular elastance (Es(sb)), arterial elastance (Ea) measured as systolic pressure • 0.9/ Stroke Volume, ventricular arterial coupling (Ea/Es(sb) ratio). Data were expressed as median and interquartile range. Measures obtained before and after the procedure were compared using Wilcoxon non parametric test for paired samples.

RESULTS:

Mitraclip procedure was effective in reducing regurgitation. We observed an amelioration of echocardiographic parameters with a reduction of estimated systolic pulmonary pressure (45 to 37,5 p = 0,0002) and left atrial volume (110 to 93 p = 0,0001). Despite a few cases decreasing in ejection fraction (37 to 35 p = 0,035), the maintained ventricular arterial coupling after the procedure (P = 0,67) was associated with an increasing in forward stroke volume (60,3 to 78 p = 0,05).

CONCLUSION:

MitraClip is effective in reducing mitral valve regurgitation and determines an amelioration of hemodynamic parameters with preservation of ventricular arterial coupling.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Próteses Valvulares Cardíacas / Ecocardiografia / Cateterismo Cardíaco / Função Ventricular Esquerda / Ventrículos do Coração / Insuficiência da Valva Mitral Tipo de estudo: Observational_studies / Risk_factors_studies Limite: Aged / Female / Humans / Male Idioma: En Revista: Cardiovasc Ultrasound Assunto da revista: ANGIOLOGIA / CARDIOLOGIA / DIAGNOSTICO POR IMAGEM Ano de publicação: 2013 Tipo de documento: Article País de afiliação: Itália

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Próteses Valvulares Cardíacas / Ecocardiografia / Cateterismo Cardíaco / Função Ventricular Esquerda / Ventrículos do Coração / Insuficiência da Valva Mitral Tipo de estudo: Observational_studies / Risk_factors_studies Limite: Aged / Female / Humans / Male Idioma: En Revista: Cardiovasc Ultrasound Assunto da revista: ANGIOLOGIA / CARDIOLOGIA / DIAGNOSTICO POR IMAGEM Ano de publicação: 2013 Tipo de documento: Article País de afiliação: Itália