Your browser doesn't support javascript.
loading
Is TKA using patient-specific instruments comparable to conventional TKA? A randomized controlled study of one system.
Roh, Yoon Whan; Kim, Tae Woo; Lee, Sahnghoon; Seong, Sang Cheol; Lee, Myung Chul.
Afiliação
  • Roh YW; Joint Reconstruction Center, Seoul NOW Hospital, Kyungki-Do, Korea.
Clin Orthop Relat Res ; 471(12): 3988-95, 2013 Dec.
Article em En | MEDLINE | ID: mdl-23907610
ABSTRACT

BACKGROUND:

Patient-specific CT-based instrumentation may reduce implant malpositioning and improve alignment in TKA. However, it is not known whether this innovation is an advance that benefits patients. QUESTIONS/

PURPOSES:

We evaluated (1) the precision of patient-specific TKA by comparing the incidence of outliers in postoperative alignment between TKAs using patient-specific instruments and TKAs using conventional instruments, and (2) the reliability of patient-specific instruments by intraoperatively investigating whether the surgery could be completed with patient-specific instruments alone.

METHODS:

In this randomized controlled trial, we compared patient-specific TKA instruments from one manufacturer (n = 50) with conventional TKA instruments (n = 50). Postoperative hip-knee-ankle angles, femoral component rotation, and coronal and sagittal alignments of each component were measured. The validity of the patient-specific instrument was examined using cross-checking procedures with conventional instruments during the surgeries. When the procedure could not be completed accurately with patient-specific instruments, the procedure was converted to TKA using conventional instruments, and the frequency of this occurrence was tallied.

RESULTS:

Outliers in the hip-knee-ankle angle were comparable between groups (12% in the patient-specific instrument group and 10% in the conventional instrument group). Other parameters such as sagittal alignment and femoral component rotation did not differ in terms of outliers. Patient-specific guides were abandoned in eight knees (16%) during the surgery because of malrotation of the femoral components and decreased slope of the tibia.

CONCLUSIONS:

Accuracy was comparable between TKAs done with patient-specific instruments and those done with conventional instruments. However, the patient-specific instrument procedures had to be aborted frequently, incurring expenses that did not benefit patients.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Artroplastia do Joelho / Osteoartrite do Joelho / Articulação do Joelho / Prótese do Joelho Tipo de estudo: Clinical_trials / Diagnostic_studies Limite: Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Clin Orthop Relat Res Ano de publicação: 2013 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Artroplastia do Joelho / Osteoartrite do Joelho / Articulação do Joelho / Prótese do Joelho Tipo de estudo: Clinical_trials / Diagnostic_studies Limite: Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Clin Orthop Relat Res Ano de publicação: 2013 Tipo de documento: Article