Your browser doesn't support javascript.
loading
Influence of pupil size on measurements made with the Lens Opacity Meter 701.
Clarke, M P; Pearson, J C; Vernon, S A; Matthews, J C.
Afiliação
  • Clarke MP; Academic Unit of Ophthalmology, Queen's Medical Centre, Nottingham.
Br J Ophthalmol ; 74(9): 526-7, 1990 Sep.
Article em En | MEDLINE | ID: mdl-2393643
ABSTRACT
The Lens Opacity Meter 701 (Interzeag, Switzerland) is a recently developed instrument for assessing cataract. It measures the degree of scatter of a red light beam by the lens. Fifty patients prior to cataract surgery had measurements made before and after dilatation of the pupil. Inconsistent results were found at pupil diameters of less than 4 mm. It is postulated that this was due to absorption of scattered light by the iris.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Oftalmologia / Catarata / Pupila Limite: Humans Idioma: En Revista: Br J Ophthalmol Ano de publicação: 1990 Tipo de documento: Article