Your browser doesn't support javascript.
loading
An institutional experience of quality assurance of a treatment planning system on photon beam.
Ozgüven, Yildiray; Yaray, Kadir; Alkaya, Fadime; Yücel, Birsen; Soyuer, Serdar.
Afiliação
  • Ozgüven Y; Department of Radiation Oncology, Cumhuriyet University, Faculty of Medicine, TR-58140 Sivas, Turkey.
  • Yaray K; Department of Radiation Oncology, Erciyes University, Faculty of Medicine, TR-38039 Kayseri, Turkey.
  • Alkaya F; Department of Radiation Oncology, Medicana Hospital, TR-34520 Istanbul, Turkey.
  • Yücel B; Department of Radiation Oncology, Cumhuriyet University, Faculty of Medicine, TR-58140 Sivas, Turkey.
  • Soyuer S; Department of Radiation Oncology, Erciyes University, Faculty of Medicine, TR-38039 Kayseri, Turkey.
Rep Pract Oncol Radiother ; 19(3): 195-205, 2014 May.
Article em En | MEDLINE | ID: mdl-24936337
ABSTRACT

AIM:

The purpose of the present study is to show the application of the IAEA TRS-430 QA procedures of Eclipse™v7.5 TPS for photon energies. In addition, the trends of the deviations found in the conducted tests were determined.

BACKGROUND:

In the past, the lack of complete TPS QA procedures led to some serious accidents. So, QA in the radiotherapy treatment planning process is essential for determination of accuracy in the radiotherapy process and avoidance of treatment errors. MATERIALS AND

METHODS:

The calculations of TPS and measurements of irradiations of the treatment device were compared in the study. As a result, the local dose deviation values (δ 1 central beam axis, δ 2 penumbra and build up region, δ 3 inside field, δ 4 outside beam edges, δ 50-90 beam fringe, RW50 radiological width) and their confidence limit values (including systematic and random errors) were obtained.

RESULTS:

The confidence limit values of δ 4 were detected to increase with expanding field size. The values of δ 1 and δ 3 of hard wedge were larger than open fields. The values of δ 2 and δ 50-90 of the inhomogeneity effect test were larger, especially than other tests of this study. The average deviation was showed to increase with the rise of the wedge angle. The values of δ 3 and δ 4 of lung irradiation were outside tolerance.

CONCLUSIONS:

The QA of TPS was done and it was found that there were no reservations in its use in patient treatment. The trend of the deviations is shown.
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Revista: Rep Pract Oncol Radiother Ano de publicação: 2014 Tipo de documento: Article País de afiliação: Turquia

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Revista: Rep Pract Oncol Radiother Ano de publicação: 2014 Tipo de documento: Article País de afiliação: Turquia