Sensitivity Enhancement of an Inductively Coupled Local Detector Using a HEMT-Based Current Amplifier.
Qian, Chunqi; Duan, Qi; Dodd, Steve; Koretsky, Alan; Murphy-Boesch, Joe.
Afiliação
  • Qian C; Laboratory of Functional and Molecular Imaging, National Institute of Neurological Disorders and Stroke, National Institutes of Health, Bethesda, Maryland, USA.
  • Duan Q; Department of Radiology, Michigan State University, East Lansing, Michigan, USA.
  • Dodd S; Laboratory of Functional and Molecular Imaging, National Institute of Neurological Disorders and Stroke, National Institutes of Health, Bethesda, Maryland, USA.
  • Koretsky A; Laboratory of Functional and Molecular Imaging, National Institute of Neurological Disorders and Stroke, National Institutes of Health, Bethesda, Maryland, USA.
  • Murphy-Boesch J; Laboratory of Functional and Molecular Imaging, National Institute of Neurological Disorders and Stroke, National Institutes of Health, Bethesda, Maryland, USA.
Magn Reson Med ; 75(6): 2573-8, 2016 06.
Article em En | MEDLINE | ID: mdl-26192998
ABSTRACT

PURPOSE:

To improve the signal transmission efficiency and sensitivity of a local detection coil that is weakly inductively coupled to a larger receive coil.

METHODS:

The resonant detection coil is connected in parallel with the gate of a high electron mobility transistor (HEMT) transistor without impedance matching. When the drain of the transistor is capacitively shunted to ground, current amplification occurs in the resonator by feedback that transforms a capacitive impedance on the transistor's source to a negative resistance on its gate.

RESULTS:

High resolution images were obtained from a mouse brain using a small, 11 mm diameter surface coil that was inductively coupled to a commercial, phased array chest coil. Although the power consumption of the amplifier was only 88 µW, 14 dB gain was obtained with excellent noise performance.

CONCLUSION:

An integrated current amplifier based on a HEMT can enhance the sensitivity of inductively coupled local detectors when weakly coupled. This amplifier enables efficient signal transmission between customized user coils and commercial clinical coils, without the need for a specialized signal interface. Magn Reson Med 752573-2578, 2016. Published 2015. This article is a U.S. Government work and is in the public domain in the USA.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Processamento de Imagem Assistida por Computador / Processamento de Sinais Assistido por Computador / Imageamento por Ressonância Magnética / Amplificadores Eletrônicos Tipo de estudo: Diagnostic_studies Limite: Animals Idioma: En Revista: Magn Reson Med Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2016 Tipo de documento: Article País de afiliação: Estados Unidos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Processamento de Imagem Assistida por Computador / Processamento de Sinais Assistido por Computador / Imageamento por Ressonância Magnética / Amplificadores Eletrônicos Tipo de estudo: Diagnostic_studies Limite: Animals Idioma: En Revista: Magn Reson Med Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2016 Tipo de documento: Article País de afiliação: Estados Unidos