Your browser doesn't support javascript.
loading
Long-Term Outcomes and Durability of the Mitroflow Aortic Bioprosthesis.
Mosquera, Victor X; Bouzas-Mosquera, Alberto; Velasco-García, Carlos; Muñiz, Javier; Estévez-Cid, Francisco; Portela-Torron, Francisco; Herrera-Noreña, José M; Cuenca-Castillo, José J.
Afiliação
  • Mosquera VX; Department of Cardiac Surgery, Complejo Hospitalario Universitario de A Coruña, A Coruña, Spain.
  • Bouzas-Mosquera A; Department of Cardiology, Complejo Hospitalario Universitario de A Coruña, A Coruña, Spain.
  • Velasco-García C; Department of Cardiac Surgery, Complejo Hospitalario Universitario de A Coruña, A Coruña, Spain.
  • Muñiz J; Instituto Universitario de Ciencias de la Salud, Universidad de A Coruña, A Coruña, Spain.
  • Estévez-Cid F; Department of Cardiac Surgery, Complejo Hospitalario Universitario de A Coruña, A Coruña, Spain.
  • Portela-Torron F; Department of Cardiac Surgery, Complejo Hospitalario Universitario de A Coruña, A Coruña, Spain.
  • Herrera-Noreña JM; Department of Cardiac Surgery, Complejo Hospitalario Universitario de A Coruña, A Coruña, Spain.
  • Cuenca-Castillo JJ; Department of Cardiac Surgery, Complejo Hospitalario Universitario de A Coruña, A Coruña, Spain.
J Card Surg ; 31(5): 264-73, 2016 May.
Article em En | MEDLINE | ID: mdl-26991888
ABSTRACT

BACKGROUND:

This study aims to determine the incidence and causes of structural valve deterioration (SVD) among all models of Mitroflow bioprostheses (A12/LX/DL), as well as to define their long-term clinical and hemodynamic performance. METHODS AND

RESULTS:

We retrospectively reviewed a series of 1023 patients who underwent aortic valve replacement with Mitroflow bioprostheses between 2001 and 2014. A small aortic root was found in 22.4% of patients. There were two cases of severe patient-prosthesis mismatch. Only 31 patients developed SVD. The rate of incidence was 8.1 cases per 1000 patient-years. Cumulative incidence of SVD was 1.4% and 3% at five and 10 years, respectively. Freedom from SVD was 97.4% and 88.2% at five and 10 years, respectively. Anticalcification phospholipid reduction treatment (PRT) for model DL was a protective factor for SVD. Multivariable analysis confirmed age <70 years and use of 19 mm valve as independent predictors of SVD. Cumulative survival was 76.6% at five years and 42.3% at 10 years (mean follow-up 3.8 ± 3.1 years). In multivariable analysis, neither the use of small aortic prosthesis (p = 0.18) nor the occurrence of SVD (p = 0.85) was found to be independent predictors of long-term survival.

CONCLUSIONS:

Mitroflow valves demonstrate an acceptable rate of SVD and satisfactory long-term hemodynamic performance, particularly in patients with small aortic roots, age >70 years, and cases with severe left ventricular hypertrophy. PRT might contribute to improved long-term durability. doi 10.1111/jocs.12726 (J Card Surg 2016;31264-273).
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Valva Aórtica / Estenose da Valva Aórtica / Complicações Pós-Operatórias / Bioprótese / Teste de Materiais / Próteses Valvulares Cardíacas / Hemodinâmica Tipo de estudo: Diagnostic_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Limite: Aged / Female / Humans / Male País/Região como assunto: Europa Idioma: En Revista: J Card Surg Assunto da revista: CARDIOLOGIA Ano de publicação: 2016 Tipo de documento: Article País de afiliação: Espanha

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Valva Aórtica / Estenose da Valva Aórtica / Complicações Pós-Operatórias / Bioprótese / Teste de Materiais / Próteses Valvulares Cardíacas / Hemodinâmica Tipo de estudo: Diagnostic_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Limite: Aged / Female / Humans / Male País/Região como assunto: Europa Idioma: En Revista: J Card Surg Assunto da revista: CARDIOLOGIA Ano de publicação: 2016 Tipo de documento: Article País de afiliação: Espanha