Your browser doesn't support javascript.
loading
A Johann-type X-ray emission spectrometer at the Rossendorf beamline.
Kvashnina, Kristina O; Scheinost, Andreas C.
Afiliação
  • Kvashnina KO; Rossendorf Beamline at ESRF - The European Synchrotron, CS40220, 38043 Grenoble Cedex 9, France.
  • Scheinost AC; Rossendorf Beamline at ESRF - The European Synchrotron, CS40220, 38043 Grenoble Cedex 9, France.
J Synchrotron Radiat ; 23(Pt 3): 836-41, 2016 05.
Article em En | MEDLINE | ID: mdl-27140166
This paper gives a detailed description, including equations, of the Johann-type X-ray emission spectrometer which has been recently installed and tested at the Rossendorf beamline (ROBL) of the European Synchrotron Radiation Facility. The spectrometer consists of a single spherically bent crystal analyzer and an avalanche photodiode detector positioned on the vertical Rowland cycle of 1 m diameter. The hard X-ray emission spectrometer (∼3.5-25 keV) operates at atmospheric pressure and covers the Bragg angles of 65°-89°. The instrument has been tested at high and intermediate incident energies, i.e. at the Zr K-edge and at the Au L3-edge, in the second experimental hutch of ROBL. The spectrometer is dedicated for studying actinides in materials and environmental samples by high-energy-resolution X-ray absorption and X-ray emission spectroscopies.
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Revista: J Synchrotron Radiat Assunto da revista: RADIOLOGIA Ano de publicação: 2016 Tipo de documento: Article País de afiliação: França

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Revista: J Synchrotron Radiat Assunto da revista: RADIOLOGIA Ano de publicação: 2016 Tipo de documento: Article País de afiliação: França