Your browser doesn't support javascript.
loading
Wound haematoma following defibrillator implantation: incidence and predictors in the Shockless Implant Evaluation (SIMPLE) trial.
Masiero, Simona; Connolly, Stuart J; Birnie, David; Neuzner, Jörg; Hohnloser, Stefan H; Vinolas, Xavier; Kautzner, Josef; O'Hara, Gilles; VanErven, Lieselot; Gadler, Fredrik; Wang, Jia; Mabo, Philippe; Glikson, Michael; Kutyifa, Valentina; Wright, David J; Essebag, Vidal; Healey, Jeff S.
Afiliação
  • Masiero S; Population Health Research Institute, McMaster University, 237 Barton Street East, Hamilton, ON, Canada L8L 2X2.
  • Connolly SJ; Clinica di Cardiologia, Università Politecnica delle Marche, via Conca 71, 60126 Ancona, Italy.
  • Birnie D; Population Health Research Institute, McMaster University, 237 Barton Street East, Hamilton, ON, CanadaL8L 2X2.
  • Neuzner J; University of Ottawa Heart Institute, 40 Ruskin Street, Ottawa, ON, CanadaK1Y 4W7.
  • Hohnloser SH; Klinikum Kassel, 43, Mönchebergstraße 41, 34125 Kassel, Germany.
  • Vinolas X; J.W. Goethe University, Theodor-W.-Adorno-Platz 6, 60323 Frankfurt am Main, Germany.
  • Kautzner J; Hospital de Santa Creu i Sant Pau, Carrer de Sant Quintí 89, 08026 Barcelona, Spain.
  • O'Hara G; Institute for Clinical and Experimental Medicine, Vídenská 1958/9, 140 21 Prague 4-Krc, Czech Republic.
  • VanErven L; Institut Universitaire de Cardiologie et de Pneumologie de Québec, 2725 Ch Ste-Foy, Québec, QC, CanadaG1V 4G5.
  • Gadler F; Leiden University Medical Center, Albinusdreef 2, 2333 ZA Leiden, The Netherlands.
  • Wang J; Karolinska Institute, Solnavägen 1, 171 77 Stockholm, Sweden.
  • Mabo P; Population Health Research Institute, McMaster University, 237 Barton Street East, Hamilton, ON, CanadaL8L 2X2.
  • Glikson M; Centre Hospitalier Universitaire, 2 Rue Henri le Guilloux, 35000 Rennes, France.
  • Kutyifa V; Leviev Heart Center, Chaim Sheba Medical Center, Tel Hashomer, 52621 Tel Aviv, Israel.
  • Wright DJ; University of Rochester Medical Center, 601 Elmwood Ave, Rochester, NY 14642, USA.
  • Essebag V; Institute of Cardiovascular Medicine and Science Liverpool Heart and Chest Hospital, Thomas Dr, Liverpool, Merseyside L14 3PE, UK.
  • Healey JS; McGill University, 845 Rue Sherbrooke O, Montrèal, QC, CanadaH3A 0G4.
Europace ; 19(6): 1002-1006, 2017 Jun 01.
Article em En | MEDLINE | ID: mdl-27353323
ABSTRACT

AIMS:

Pocket haematoma is a common complication after defibrillator [implantable cardioverter defibrillator (ICD)] implantation, which is not only painful, but also increases the risk of device-related infection, and possibly embolic events. The present study seeks to evaluate the rate and predictors of clinically significant pocket haematoma. METHODS AND

RESULTS:

This study included 2500 patients receiving an ICD in the SIMPLE trial. A clinically significant pocket haematoma was defined as a haematoma that required re-operation or interruption of oral anticoagulation (OAC) therapy. Clinically significant pocket haematoma occurred in 56 of 2500 patients (2.2%) of which 6 (10.7%) developed device-related infection. Patients who developed pocket haematoma were older (mean age 67.6 ± 8.8 years vs. 62.7 ± 11.6 years, P < 0.001), were more likely to have permanent atrial fibrillation (30.4 vs. 6.7%, P < 0.001) and a history of stroke (17.9 vs. 6.7%, P = 0.004), or were more likely to receive peri-operative OAC (50.0 vs. 28.4%, P < 0.001), unfractionated heparin (16.1 vs. 5.2%, P = 0.003), or low-molecular-weight heparin (37.5 vs. 17.5%, P < 0.001). Independent predictors of wound haematoma on multivariable analysis included the use of heparin bridging (OR 2.65, 95% CI 1.48-4.73, P = 0.001), sub-pectoral location of ICD (OR 2.00, 95% CI 1.12-3.57, P =0.020), previous stroke (OR 2.47, 95% CI 1.20-5.10, P = 0.015), an upgrade from permanent pacemaker (OR 2.52, 95% CI 1.07-5.94, P = 0.035), and older age (OR 1.03, 95% CI 1.00-1.06, P = 0.049).

CONCLUSION:

Pocket haematoma remains an important complication of ICD implantation and is associated with a high risk of infection. Independent predictors of pocket haematoma include heparin bridging, prior stroke, sub-pectoral placement of ICD, older age, and upgrade from a pacemaker.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Arritmias Cardíacas / Cardioversão Elétrica / Desfibriladores Implantáveis / Implantação de Prótese / Ferida Cirúrgica / Hematoma Tipo de estudo: Clinical_trials / Diagnostic_studies / Etiology_studies / Incidence_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Limite: Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Europace Assunto da revista: CARDIOLOGIA / FISIOLOGIA Ano de publicação: 2017 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Arritmias Cardíacas / Cardioversão Elétrica / Desfibriladores Implantáveis / Implantação de Prótese / Ferida Cirúrgica / Hematoma Tipo de estudo: Clinical_trials / Diagnostic_studies / Etiology_studies / Incidence_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Limite: Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Europace Assunto da revista: CARDIOLOGIA / FISIOLOGIA Ano de publicação: 2017 Tipo de documento: Article