Comparing an accelerated 3D fast spin-echo sequence (CS-SPACE) for knee 3-T magnetic resonance imaging with traditional 3D fast spin-echo (SPACE) and routine 2D sequences.
Altahawi, Faysal F; Blount, Kevin J; Morley, Nicholas P; Raithel, Esther; Omar, Imran M.
Afiliação
  • Altahawi FF; Department of Radiology, Northwestern University Feinberg School of Medicine, 676 N Saint Clair St Suite 800, Chicago, IL, 60611, USA. faysalaltahawi@gmail.com.
  • Blount KJ; Department of Radiology, Northwestern University Feinberg School of Medicine, 676 N Saint Clair St Suite 800, Chicago, IL, 60611, USA.
  • Morley NP; Department of Radiology, Marshfield Clinic, Marshfield, WI, USA.
  • Raithel E; Siemens Healthcare GmbH, Erlangen, Germany.
  • Omar IM; Department of Radiology, Northwestern University Feinberg School of Medicine, 676 N Saint Clair St Suite 800, Chicago, IL, 60611, USA.
Skeletal Radiol ; 46(1): 7-15, 2017 Jan.
Article em En | MEDLINE | ID: mdl-27744578
ABSTRACT

PURPOSE:

To compare a faster, new, high-resolution accelerated 3D-fast-spin-echo (3D-FSE) acquisition sequence (CS-SPACE) to traditional 2D and high-resolution 3D sequences for knee 3-T magnetic resonance imaging (MRI). MATERIALS AND

METHODS:

Twenty patients received knee MRIs that included routine 2D (T1, PD ± FS, T2-FS; 0.5 × 0.5 × 3 mm3; ∼10 min), traditional 3D FSE (SPACE-PD-FS; 0.5 × 0.5 × 0.5 mm3; ∼7.5 min), and accelerated 3D-FSE prototype (CS-SPACE-PD-FS; 0.5 × 0.5 × 0.5 mm3; ∼5 min) acquisitions on a 3-T MRI system (Siemens MAGNETOM Skyra). Three musculoskeletal radiologists (MSKRs) prospectively and independently reviewed the studies with graded surveys comparing image and diagnostic quality. Tissue-specific signal-to-noise ratios (SNR) and contrast-to-noise ratios (CNR) were also compared.

RESULTS:

MSKR-perceived diagnostic quality of cartilage was significantly higher for CS-SPACE than for SPACE and 2D sequences (p < 0.001). Assessment of diagnostic quality of menisci and synovial fluid was higher for CS-SPACE than for SPACE (p < 0.001). CS-SPACE was not significantly different from SPACE but had lower assessments than 2D sequences for evaluation of bones, ligaments, muscles, and fat (p ≤ 0.004). 3D sequences had higher spatial resolution, but lower overall assessed contrast (p < 0.001). Overall image quality from CS-SPACE was assessed as higher than SPACE (p = 0.007), but lower than 2D sequences (p < 0.001). Compared to SPACE, CS-SPACE had higher fluid SNR and CNR against all other tissues (all p < 0.001).

CONCLUSIONS:

The CS-SPACE prototype allows for faster isotropic acquisitions of knee MRIs over currently used protocols. High fluid-to-cartilage CNR and higher spatial resolution over routine 2D sequences may present a valuable role for CS-SPACE in the evaluation of cartilage and menisci.
Assuntos
Palavras-chave
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Processamento de Imagem Assistida por Computador / Imageamento por Ressonância Magnética / Aumento da Imagem / Imageamento Tridimensional / Articulação do Joelho Tipo de estudo: Guideline / Observational_studies Limite: Adolescent / Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Skeletal Radiol Ano de publicação: 2017 Tipo de documento: Article País de afiliação: Estados Unidos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Processamento de Imagem Assistida por Computador / Imageamento por Ressonância Magnética / Aumento da Imagem / Imageamento Tridimensional / Articulação do Joelho Tipo de estudo: Guideline / Observational_studies Limite: Adolescent / Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Skeletal Radiol Ano de publicação: 2017 Tipo de documento: Article País de afiliação: Estados Unidos