Usability and reproducibility of tear meniscus values generated via swept-source optical coherence tomography and the slit lamp with a graticule method.
Imamura, Hitoshi; Tabuchi, Hitoshi; Nakakura, Shunsuke; Nagasato, Daisuke; Baba, Hiroaki; Kiuchi, Yoshiaki.
Afiliação
  • Imamura H; Department of Ophthalmology, Saneikai Tsukazaki Hospital, 68-1, Aboshi Waku, Himeji, 671-1227, Japan. h.imamura@tsukazaki-eye.net.
  • Tabuchi H; Department of Ophthalmology and Visual Sciences, Graduate School of Biomedical Sciences, Hiroshima University, Hiroshima, Japan. h.imamura@tsukazaki-eye.net.
  • Nakakura S; Department of Ophthalmology, Saneikai Tsukazaki Hospital, 68-1, Aboshi Waku, Himeji, 671-1227, Japan.
  • Nagasato D; Department of Ophthalmology, Saneikai Tsukazaki Hospital, 68-1, Aboshi Waku, Himeji, 671-1227, Japan.
  • Baba H; Department of Ophthalmology, Saneikai Tsukazaki Hospital, 68-1, Aboshi Waku, Himeji, 671-1227, Japan.
  • Kiuchi Y; Department of Ophthalmology, Saneikai Tsukazaki Hospital, 68-1, Aboshi Waku, Himeji, 671-1227, Japan.
Int Ophthalmol ; 38(2): 679-686, 2018 Apr.
Article em En | MEDLINE | ID: mdl-28393321
ABSTRACT

PURPOSE:

To investigate the usability and the reproducibility of the tear meniscus values via swept-source optical coherence tomography (SS-OCT) and the conventional slit lamp microscope method with a graticule.

METHODS:

The right eye was examined in 90 healthy adult subjects who were grouped according to age (group 1 20-39 years; group 2 40-59 years; group 3 ≥60 years). The tear meniscus height (TMH) and tear meniscus area were measured using SS-OCT and TMH by the slit lamp microscope method. The reproducibility of each method was calculated using intraclass correlation coefficients (ICCs) in additionally enrolled 30 healthy young subjects. We also evaluated TMH at 3 mm from the corneal center in both temporal and nasal directions using SS-OCT.

RESULTS:

The mean of the TMH values measured by SS-OCT was significantly higher than those measured by the slit lamp method (328 vs. 212 µm, P < 0.001, respectively). High reproducibility was observed for each method (ICC > 0.75 for both). No statistically significant differences were found in TMH among the age groups using both SS-OCT and slit lamp methods (P = 0.985, 0.380, respectively). TMH values at both sides of the corneal center were significantly smaller than those at the corneal center (P < 0.0001).

CONCLUSIONS:

TMH values obtained by the slit lamp method were lower than those obtained by SS-OCT. However, both methods yielded highly reproducible TMH measurements, suggesting that they are clinically useful. Tear meniscus values did not vary by age but by measurement points in our cohort.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Lágrimas / Tomografia de Coerência Óptica / Microscopia com Lâmpada de Fenda Tipo de estudo: Diagnostic_studies Limite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Int Ophthalmol Ano de publicação: 2018 Tipo de documento: Article País de afiliação: Japão

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Lágrimas / Tomografia de Coerência Óptica / Microscopia com Lâmpada de Fenda Tipo de estudo: Diagnostic_studies Limite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Int Ophthalmol Ano de publicação: 2018 Tipo de documento: Article País de afiliação: Japão