Your browser doesn't support javascript.
loading
3D diffusion-weighted 129 Xe MRI for whole lung morphometry.
Chan, Ho-Fung; Stewart, Neil J; Norquay, Graham; Collier, Guilhem J; Wild, Jim M.
Afiliação
  • Chan HF; POLARIS, Academic Unit of Radiology, University of Sheffield, Sheffield, UK.
  • Stewart NJ; POLARIS, Academic Unit of Radiology, University of Sheffield, Sheffield, UK.
  • Norquay G; POLARIS, Academic Unit of Radiology, University of Sheffield, Sheffield, UK.
  • Collier GJ; POLARIS, Academic Unit of Radiology, University of Sheffield, Sheffield, UK.
  • Wild JM; POLARIS, Academic Unit of Radiology, University of Sheffield, Sheffield, UK.
Magn Reson Med ; 79(6): 2986-2995, 2018 06.
Article em En | MEDLINE | ID: mdl-29034509
ABSTRACT

PURPOSE:

To obtain whole lung morphometry measurements from 129 Xe in a single breath-hold with 3D multiple b-value 129 Xe diffusion-weighted MRI (DW-MRI) with an empirically optimized diffusion time and compressed sensing for scan acceleration.

METHODS:

Prospective three-fold undersampled 3D multiple b-value hyperpolarized 129 Xe DW-MRI datasets were acquired, and the diffusion time (Δ) was iterated so as to provide diffusive length scale (LmD ) estimates from the stretched exponential model (SEM) that are comparable to those from 3 He. The empirically optimized 129 Xe diffusion time was then implemented with a four-fold undersampling scheme and was prospectively benchmarked against 3 He measurements in a cohort of five healthy volunteers, six ex-smokers, and two chronic obstructive pulmonary disease patients using both SEM-derived LmD and cylinder model (CM)-derived mean chord length (Lm).

RESULTS:

Good agreement between the mean 129 Xe and 3 He LmD (mean difference, 2.2%) and Lm (mean difference, 1.1%) values was obtained in all subjects at an empirically optimized 129 Xe Δ = 8.5 ms.

CONCLUSION:

Compressed sensing has facilitated single-breath 3D multiple b-value 129 Xe DW-MRI acquisitions, and results at 129 Xe Δ = 8.5 ms indicate that 129 Xe provides a viable alternative to 3 He for whole lung morphometry mapping with either the SEM or CM. Magn Reson Med 792986-2995, 2018. © 2017 The Authors Magnetic Resonance in Medicine published by Wiley Periodicals, Inc. on behalf of International Society for Magnetic Resonance in Medicine. This is an open access article under the terms of the Creative Commons Attribution License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Isótopos de Xenônio / Imageamento Tridimensional / Doença Pulmonar Obstrutiva Crônica / Imagem de Difusão por Ressonância Magnética / Pulmão Tipo de estudo: Etiology_studies / Incidence_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Limite: Adult / Female / Humans / Male / Middle aged Idioma: En Revista: Magn Reson Med Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2018 Tipo de documento: Article País de afiliação: Reino Unido

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Isótopos de Xenônio / Imageamento Tridimensional / Doença Pulmonar Obstrutiva Crônica / Imagem de Difusão por Ressonância Magnética / Pulmão Tipo de estudo: Etiology_studies / Incidence_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Limite: Adult / Female / Humans / Male / Middle aged Idioma: En Revista: Magn Reson Med Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2018 Tipo de documento: Article País de afiliação: Reino Unido