Investigating the effect of coil model losses on computational electromagnetic exposure of an ASTM phantom at 64 MHz MRI.
Annu Int Conf IEEE Eng Med Biol Soc ; 2017: 1481-1484, 2017 Jul.
Article em En | MEDLINE | ID: mdl-29060159
ABSTRACT
The goal of this work is to investigate the effect of coil losses on the electromagnetic field generated in an ASTM phantom by a birdcage coil. The study was based on different numerical implementations of an RF body coil at 64 MHz, using the same 3D EM and RF circuit co-simulation procedure. The coil quality factor was evaluated with respect to losses due to power feed mismatch and to resistive losses of the coil components. The results of the study showed that the magnetic field at the coil iso-center, normalized to the square root of the whole body specific absorption rate, depends on the coil quality factor.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Imagens de Fantasmas Tipo de estudo: Prognostic_studies Idioma: En Revista: Annu Int Conf IEEE Eng Med Biol Soc Ano de publicação: 2017 Tipo de documento: Article