Your browser doesn't support javascript.
loading
Flow-volume loops measured with electrical impedance tomography in pediatric patients with asthma.
Ngo, Chuong; Dippel, Falk; Tenbrock, Klaus; Leonhardt, Steffen; Lehmann, Sylvia.
Afiliação
  • Ngo C; Philips Chair for Medical Information Technology, RWTH Aachen, Aachen, Germany.
  • Dippel F; Philips Chair for Medical Information Technology, RWTH Aachen, Aachen, Germany.
  • Tenbrock K; Department of Pediatric Pulmonology, RWTH Aachen University Hospital, Aachen, Germany.
  • Leonhardt S; Philips Chair for Medical Information Technology, RWTH Aachen, Aachen, Germany.
  • Lehmann S; Department of Pediatric Pulmonology, RWTH Aachen University Hospital, Aachen, Germany.
Pediatr Pulmonol ; 53(5): 636-644, 2018 05.
Article em En | MEDLINE | ID: mdl-29405616
ABSTRACT

BACKGROUND:

Electrical impedance tomography (EIT) provides information on global and regional ventilation during tidal breathing and mechanical ventilation. During forced expiration maneuvers, the linearity of EIT and spirometric data has been documented in healthy persons. The present study investigates the potential diagnostic use of EIT in pediatric patients with asthma.

METHODS:

EIT and spirometry were performed in 58 children with asthma (average age ± SD 11.86 ± 3.13 years), and 58 healthy controls (average age ± SD 12.12 ± 2.9 years). The correlation between EIT data and simultaneously acquired spirometric data were tested for FEV1, FEV0.5 , MEF75 , MEF50 , and MEF25 . Binary classification tests were performed for the EIT-derived Tiffeneau index FEV1 /FVC and the bronchodilator test index ΔFEV1 . Average flow-volume (FV) loops were generated for patients with pathologic spirometry to demonstrate the feasibility of EIT for graphic diagnosis of asthma.

RESULTS:

Spirometry and global EIT-based FV loops showed a strong correlation (P < 0.001, r > 0.9 in FEV1 and FEV0.5 ). In all criteria, the binary classification tests yielded high specificity (>93%), a high positive predictive value (≥75%) and a high negative predictive value (>80%), while sensitivity was higher in ΔFEV1 (86.67%) and lower in FEV1 /FVC (25% and 35.29%). A typical concave shape of the EIT-derived average FV loops was observed for asthmatic children with improvement after bronchospasmolysis.

CONCLUSIONS:

Global FV loops derived from EIT correlate well with spirometry. Positive bronchospasmolysis can be observed in EIT-derived FV loops. Flow-volume loops originated from EIT have a potential to visualize pulmonary function.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Asma / Tomografia / Impedância Elétrica Tipo de estudo: Diagnostic_studies / Prognostic_studies Limite: Adolescent / Child / Female / Humans / Male Idioma: En Revista: Pediatr Pulmonol Assunto da revista: PEDIATRIA Ano de publicação: 2018 Tipo de documento: Article País de afiliação: Alemanha

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Asma / Tomografia / Impedância Elétrica Tipo de estudo: Diagnostic_studies / Prognostic_studies Limite: Adolescent / Child / Female / Humans / Male Idioma: En Revista: Pediatr Pulmonol Assunto da revista: PEDIATRIA Ano de publicação: 2018 Tipo de documento: Article País de afiliação: Alemanha