Evaluation of Four Blood Glucose Monitoring Systems for Self-Testing with Built-in Insulin Dose Advisor Based on ISO 15197:2013: System Accuracy and Hematocrit Influence.
Jendrike, Nina; Baumstark, Annette; Pleus, Stefan; Liebing, Christina; Beer, Alexandra; Flacke, Frank; Haug, Cornelia; Freckmann, Guido.
Afiliação
  • Jendrike N; 1 Institut für Diabetes-Technologie Forschungs- und Entwicklungsgesellschaft mbH an der Universität Ulm , Ulm, Germany .
  • Baumstark A; 1 Institut für Diabetes-Technologie Forschungs- und Entwicklungsgesellschaft mbH an der Universität Ulm , Ulm, Germany .
  • Pleus S; 1 Institut für Diabetes-Technologie Forschungs- und Entwicklungsgesellschaft mbH an der Universität Ulm , Ulm, Germany .
  • Liebing C; 1 Institut für Diabetes-Technologie Forschungs- und Entwicklungsgesellschaft mbH an der Universität Ulm , Ulm, Germany .
  • Beer A; 2 Sanofi-Aventis Deutschland GmbH, Industriepark Höchst , Frankfurt am Main, Germany .
  • Flacke F; 2 Sanofi-Aventis Deutschland GmbH, Industriepark Höchst , Frankfurt am Main, Germany .
  • Haug C; 1 Institut für Diabetes-Technologie Forschungs- und Entwicklungsgesellschaft mbH an der Universität Ulm , Ulm, Germany .
  • Freckmann G; 1 Institut für Diabetes-Technologie Forschungs- und Entwicklungsgesellschaft mbH an der Universität Ulm , Ulm, Germany .
Diabetes Technol Ther ; 20(4): 303-313, 2018 04.
Article em En | MEDLINE | ID: mdl-29664706
BACKGROUND: Self-monitoring of blood glucose (SMBG) is important in diabetes therapy; however, not all SMBG systems are sufficiently accurate. In addition, some SMBG systems are influenced by the user's hematocrit value. METHODS: System accuracy and hematocrit influence was evaluated for four SMBG systems with built-in insulin dose advisors (Accu-Chek® Aviva Expert [1], FreeStyle InsuLinx [2], FreeStyle Precision Neo [3], MyStar DoseCoach® [4]) based on International Organization for Standardization (ISO) 15197:2013 section 6.3 (system accuracy) and 6.4.3 (packed cell volume [hematocrit]) with three test strip lots for each system. Two different established comparison methods were used to investigate a possible impact of the comparison method on analytical performance data. RESULTS: Two systems (2, 4) fulfilled ISO 15197:2013 accuracy criteria when the manufacturer's comparison measurement method was applied and showed with all three tested lots 97% to 99.5% of results within ±15 mg/dL and ±15% of the comparison measurement results at blood glucose (BG) concentrations <100 and ≥100 mg/dL, respectively, and 100% of results within consensus error grid zones A and B. Regarding hematocrit influences, two systems (3, 4) showed with all three tested lots ≤10 mg/dL and ≤10% difference between the test sample and the respective control sample for BG concentrations <100 and ≥100 mg/dL, respectively, when using the manufacturer's comparison measurement method. CONCLUSIONS: When using the manufacturer's comparison measurement method, two out of four SMBG systems fulfilled the minimum system accuracy requirements of ISO 15197:2013. In addition, varying hematocrit levels can affect measurement results with some SMBG systems with built-in insulin dose advisors.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Glicemia / Sistemas de Infusão de Insulina / Automonitorização da Glicemia / Diabetes Mellitus Tipo de estudo: Evaluation_studies Limite: Adolescent / Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Diabetes Technol Ther Assunto da revista: ENDOCRINOLOGIA / TERAPEUTICA Ano de publicação: 2018 Tipo de documento: Article País de afiliação: Alemanha

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Glicemia / Sistemas de Infusão de Insulina / Automonitorização da Glicemia / Diabetes Mellitus Tipo de estudo: Evaluation_studies Limite: Adolescent / Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Diabetes Technol Ther Assunto da revista: ENDOCRINOLOGIA / TERAPEUTICA Ano de publicação: 2018 Tipo de documento: Article País de afiliação: Alemanha