Wear Kinetics of Highly Cross-Linked and Conventional Polyethylene Are Similar at Medium-term Follow-Up After Primary Total Hip Arthroplasty.
Mayer, Constantin; Rommelmann, Moritz; Behringer, Michael; Jäger, Marcus; Krauspe, Rüdiger; Zilkens, Christoph.
Afiliação
  • Mayer C; Department of Orthopedics, Heinrich Heine University Medical School, Duesseldorf, Germany; Department of Orthopedics and Trauma Surgery, University Medical School, Essen, Germany.
  • Rommelmann M; Department of Orthopedics, Heinrich Heine University Medical School, Duesseldorf, Germany.
  • Behringer M; Institute of Sports Sciences, Goethe University Frankfurt, Frankfurt, Germany.
  • Jäger M; Department of Orthopedics and Trauma Surgery, University Medical School, Essen, Germany.
  • Krauspe R; Department of Orthopedics, Heinrich Heine University Medical School, Duesseldorf, Germany.
  • Zilkens C; Department of Orthopedics, Heinrich Heine University Medical School, Duesseldorf, Germany.
J Arthroplasty ; 33(8): 2671-2676, 2018 08.
Article em En | MEDLINE | ID: mdl-29699828
ABSTRACT

BACKGROUND:

Wear debris is a major factor in aseptic loosening of total hip arthroplasty. Ultra high molecular weight polyethylene inlays are known for significant wear, and the following generation, highly cross-linked polyethylene (HCLPE), has shown promising in vitro and short-term in vivo results. This study aimed to investigate wear debris of HCLPE liners with ceramic heads after 9 years to reveal the in vivo wear kinetics of this common bearing combination.

METHODS:

Fifty-seven patients (72 hips; 46.5 ± 15.5 years; range 16-76 years) who underwent hip arthroplasty with an HCLPE liner (28- or 32-mm Biolox forte ceramic head) were followed up (mean 9.1 ± 2.4 years; range 3.9-13.8 years). Conventional anteroposterior X-rays were analyzed using Hip Analysis Suite software.

RESULTS:

Volumetric wear had a mean of 38.67 ± 22.09 mm3/year, 333.08 ± 183.93 mm3 overall, and linear wear was 0.063 ± 0.03 mm/year and 0.546 ± 0.27 mm overall. Male patients had a significantly higher wear rate (46.42 ± 27.68 mm3/year) and total wear (400.71 ± 235.21 mm3). Larger femoral heads had a significantly higher wear rate (43.10 ± 23.93 mm3/year) and total wear (364.23 ± 203.68 mm3). Regression analysis showed a significant cubic relationship (R2 = 0.307) with increasing yearly wear after approximately 108 months postoperatively.

CONCLUSIONS:

HCLPE liners show significant in vivo wear after 9 years. While the total wear compared to ultra high molecular weight polyethylene liners was decreased, the wear kinetics show a comparable course. The increase in wear rate after only 108 months postoperatively is especially alarming. Longer term follow-up is needed to distinguish the long-term superiority of HCLPE liners in polyethylene-ceramic paired hip arthroplasty.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Polietilenos / Falha de Prótese / Artroplastia de Quadril / Polietileno Tipo de estudo: Diagnostic_studies / Observational_studies / Prognostic_studies Limite: Adolescent / Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: J Arthroplasty Assunto da revista: ORTOPEDIA Ano de publicação: 2018 Tipo de documento: Article País de afiliação: Alemanha

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Polietilenos / Falha de Prótese / Artroplastia de Quadril / Polietileno Tipo de estudo: Diagnostic_studies / Observational_studies / Prognostic_studies Limite: Adolescent / Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: J Arthroplasty Assunto da revista: ORTOPEDIA Ano de publicação: 2018 Tipo de documento: Article País de afiliação: Alemanha