Fourier transform infrared spectrometer based on an electrothermal MEMS mirror.
Appl Opt ; 57(21): 5956-5961, 2018 Jul 20.
Article em En | MEDLINE | ID: mdl-30118019
ABSTRACT
This paper presents a micro Fourier transform infrared spectrometer (µFTIR), enabled by an H-shaped electrothermal microelectromechanical systems (MEMS) mirror. A special driving method was developed for obtaining a linear, uniform-speed motion of 186 µm, and the tilting angle of the MEMS mirror was as small as 0.06°, so there was no need of complex closed-loop control. A telecentric lens was employed in the interferometer of the µFTIR to reduce the influence of the MEMS mirror tilting effect. Also, a new phase interpolation algorithm, instead of the traditional fringe interval method, was applied in the process of the spectral reconstruction to improve the spectral stability. Finally, the new µFTIR was applied in the composition prediction of soybeans, and the experimental results show that it can accurately measure grain moisture, protein, and fat contents.

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Revista: Appl Opt Ano de publicação: 2018 Tipo de documento: Article