Your browser doesn't support javascript.
loading
Comparison of manual and dose reduction modes of a MORITA R100 CBCT.
Rottke, Dennis; Dreger, Julia; Sawada, Kunihiko; Honda, Kazuya; Schulze, Dirk.
Afiliação
  • Rottke D; 1 Digital Diagnostic Center , Freiburg im Breisgau , Germany.
  • Dreger J; 2 Department of Odontology, Oral and Maxillofacial Radiology, Faculty of Medicine, Umeå University , Umeå , Sweden.
  • Sawada K; 1 Digital Diagnostic Center , Freiburg im Breisgau , Germany.
  • Honda K; 3 Department of Radiology, Nihon University School of Dentistry , Tokyo , Japan.
  • Schulze D; 3 Department of Radiology, Nihon University School of Dentistry , Tokyo , Japan.
Dentomaxillofac Radiol ; 48(2): 20180009, 2019 Feb.
Article em En | MEDLINE | ID: mdl-30182744
ABSTRACT

OBJECTIVE:

To compare effective dose differences when acquiring (1) dose reduction mode and (2) manual mode in a MORITA R100 CBCT.

METHODS:

24 exposure protocols with different technique factors were performed in both the dose reduction mode and the manual mode in a Veraviewepocs 3D R100 cone beam CT device. 48 TLD were placed in a RANDO head phantom at 24 different sites. Effective doses were calculated according to the formalism published in the 103rd ICRP report.

RESULTS:

Effective doses for the dose reduction mode protocols ranged from 14 to 156 µSv [mean = 57 µSv, standard deviation (SD) = 37 µSv], whereas effective doses for the manual mode protocols ranged from 22 to 267 µSv (mean = 94 µSv, SD = 65 µSv). Thus, across all protocols, the dose reduction mode leads to a drop of the effective dose by 38 % (SD = 6 %).

CONCLUSIONS:

The estimated effective doses are significantly lower if dose reduction protocols are acquired.