Comparison of two noninvasive techniques for estimating cardiac output during exercise.
J Appl Physiol (1985)
; 61(1): 155-9, 1986 Jul.
Article
em En
| MEDLINE
| ID: mdl-3090011
This study presents the comparison of two different noninvasive techniques for the estimation of cardiac output (Q). The two techniques used were transthoracic impedance plethysmography (Z) and the indirect Fick CO2 rebreathing (RB) method. Paired estimates of Q were made on 60 different male subjects at rest and during graded increments of work on a cycle ergometer. The mean resting Q as measured by the Z technique (COZ) was 7.46 +/- 0.35 and 5.96 +/- 0.43 l/min using the RB (CORB) technique. At 200 W the mean COZ was 18.67 +/- 0.72 l/min and the CORB was 23.73 +/- 0.84 l/min. Both the techniques were linearly correlated (R) with O2 consumption; i.e., RZ = 0.752, RRB = 0.855. The difference between these two R values is statistically significant (P less than 0.001). A linear relationship was found between the Z and RB techniques at all work loads (R = 0.75). This study suggests that both techniques are equally as reliable over a large range of work loads, with the Z technique being the simplest and most efficient to implement. It was also found that lung volume had no effect on the calculated COZ.