Your browser doesn't support javascript.
loading
Pacemaker and Defibrillator Implantation and Programming in Patients with Deep Brain Stimulation.
Elliott, Mark; Momin, Sheikh; Fiddes, Barnaby; Farooqi, Fahad; Sohaib, Sm Afzal.
Afiliação
  • Elliott M; Department of Cardiology, King George Hospital Ilford, UK.
  • Momin S; Department of Cardiology, King George Hospital Ilford, UK.
  • Fiddes B; Barking, Havering and Redbridge University Hospitals NHS Trust Romford, UK.
  • Farooqi F; Department of Cardiology, King George Hospital Ilford, UK.
  • Sohaib SA; Department of Cardiology, King George Hospital Ilford, UK.
Arrhythm Electrophysiol Rev ; 8(2): 138-142, 2019 05.
Article em En | MEDLINE | ID: mdl-31114689
ABSTRACT
The need for cardiac device implantation in patients receiving deep brain stimulation (DBS) is increasing. Despite the theoretical risk of the two systems interacting, there are no clear guidelines for cardiologists carrying out cardiac device implantation in this population. We performed a review of the literature and describe 13 case reports in which patients have both DBS and a cardiac pacemaker or ICD implanted. Except for one early study, in which an ICD shock reset the deep brain stimulator, no significant interactions have been reported. We discuss the potential interactions between DBS and cardiac devices, and provide practical advice for implanting cardiologists. We conclude that, provided that specific precautions are taken, cardiac device implantation is likely to be safe in patients with DBS.
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Tipo de estudo: Guideline Idioma: En Revista: Arrhythm Electrophysiol Rev Ano de publicação: 2019 Tipo de documento: Article País de afiliação: Reino Unido

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Tipo de estudo: Guideline Idioma: En Revista: Arrhythm Electrophysiol Rev Ano de publicação: 2019 Tipo de documento: Article País de afiliação: Reino Unido