Your browser doesn't support javascript.
loading
How to deliver personalized cardiac resynchronization therapy through the precise measurement of the acute hemodynamic response: Insights from the iSpot trial.
Shun-Shin, Matthew J; Miyazawa, Alejandra A; Keene, Daniel; Sterlinski, Maciej; Sokal, Adam; Van Heuverswyn, Frédéric; Rinaldi, Christopher A; Cornelussen, Richard; Stegemann, Berthold; Francis, Darrel P; Whinnett, Zachary.
Afiliação
  • Shun-Shin MJ; International Centre for Circulatory Health, Imperial College London, Hammersmith Hospital, London, United Kingdom.
  • Miyazawa AA; International Centre for Circulatory Health, Imperial College London, Hammersmith Hospital, London, United Kingdom.
  • Keene D; International Centre for Circulatory Health, Imperial College London, Hammersmith Hospital, London, United Kingdom.
  • Sterlinski M; The Second Department of Coronary Artery Disease, Institute of Cardiology, Warsaw, Poland.
  • Sokal A; Department of Cardiology, Congenital Heart Diseases and Electrotherapy, Silesian Center of Heart Disease, Zabrze, Poland.
  • Van Heuverswyn F; Heart Center, Ghent University Hospital, Ghent, Belgium.
  • Rinaldi CA; St Thomas' Hospital, Guy's and St Thomas' NHS Trust, London, England.
  • Cornelussen R; Bakken Research Center B.V., Research and Technology, Maastricht, The Netherlands.
  • Stegemann B; Bakken Research Center B.V., Research and Technology, Maastricht, The Netherlands.
  • Francis DP; International Centre for Circulatory Health, Imperial College London, Hammersmith Hospital, London, United Kingdom.
  • Whinnett Z; International Centre for Circulatory Health, Imperial College London, Hammersmith Hospital, London, United Kingdom.
J Cardiovasc Electrophysiol ; 30(9): 1610-1619, 2019 09.
Article em En | MEDLINE | ID: mdl-31115945
ABSTRACT

INTRODUCTION:

New pacing technologies offer a greater choice of left ventricular pacing sites and greater personalization of cardiac resynchronization therapy (CRT). The effects on cardiac function of novel pacing configurations are often compared using multi-beat averages of acute hemodynamic measurements. In this analysis of the iSpot trial, we explore whether this is sufficient. MATERIALS AND

METHODS:

The iSpot trial was an international, prospective, acute hemodynamic trial that assessed seven CRT configurations standard CRT, MultiSpot (posterolateral vein), and MultiVein (anterior and posterior vein) pacing. Invasive and noninvasive blood pressure, and left ventricular (LV) dP/dtmax were recorded. Eight beats were recorded before and after an alternation from AAI to the tested pacing configuration and vice-versa. Eight alternations were performed for each configuration at each of the five atrioventricular delays.

RESULTS:

Twenty-five patients underwent the full protocol of eight alternations. Only four (16%) patients had a statistically significant >3 mm Hg improvement over conventional CRT configuration (posterolateral vein, distal electrode). However, if only one alternation was analyzed (standard multi-beat averaging protocol), 15 (60%) patients falsely appeared to have a superior nonconventional configuration. Responses to pacing were significantly correlated between the different hemodynamic

measures:

invasive systolic blood pressure (SBP) vs noninvasive SBP r = 0.82 (P < .001); invasive SBP vs LV dP/dt r = 0.57, r2 = 0.32 (P < .001).

CONCLUSIONS:

Current standard multibeat acquisition protocols are unfortunately unable to prevent false impressions of optimality arising in individual patients. Personalization processes need to include distinct repeated transitions to the tested pacing configuration in addition to averaging multiple beats. The need is not only during research stages but also during clinical implementation.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Bloqueio de Ramo / Terapia de Ressincronização Cardíaca / Monitorização Hemodinâmica / Insuficiência Cardíaca / Hemodinâmica Tipo de estudo: Clinical_trials / Diagnostic_studies / Guideline / Observational_studies / Prognostic_studies / Risk_factors_studies Limite: Humans Idioma: En Revista: J Cardiovasc Electrophysiol Assunto da revista: ANGIOLOGIA / CARDIOLOGIA / FISIOLOGIA Ano de publicação: 2019 Tipo de documento: Article País de afiliação: Reino Unido

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Bloqueio de Ramo / Terapia de Ressincronização Cardíaca / Monitorização Hemodinâmica / Insuficiência Cardíaca / Hemodinâmica Tipo de estudo: Clinical_trials / Diagnostic_studies / Guideline / Observational_studies / Prognostic_studies / Risk_factors_studies Limite: Humans Idioma: En Revista: J Cardiovasc Electrophysiol Assunto da revista: ANGIOLOGIA / CARDIOLOGIA / FISIOLOGIA Ano de publicação: 2019 Tipo de documento: Article País de afiliação: Reino Unido