Prospective motion and B0 shim correction for MR spectroscopy in human brain at 7T.
Deelchand, Dinesh K; Joers, James M; Auerbach, Edward J; Henry, Pierre-Gilles.
Afiliação
  • Deelchand DK; Center for Magnetic Resonance Research and Department of Radiology, University of Minnesota, Minneapolis, Minnesota.
  • Joers JM; Center for Magnetic Resonance Research and Department of Radiology, University of Minnesota, Minneapolis, Minnesota.
  • Auerbach EJ; Center for Magnetic Resonance Research and Department of Radiology, University of Minnesota, Minneapolis, Minnesota.
  • Henry PG; Center for Magnetic Resonance Research and Department of Radiology, University of Minnesota, Minneapolis, Minnesota.
Magn Reson Med ; 82(6): 1984-1992, 2019 12.
Article em En | MEDLINE | ID: mdl-31297889
ABSTRACT

PURPOSE:

To demonstrate feasibility and performance of prospective motion and B0 shim correction for MRS in human brain at 7T.

METHODS:

Prospective motion correction using an optical camera and linear B0 shim correction using FASTMAP-like navigators were implemented into a semi-LASER sequence. The effect of motion on spectral quality was assessed without and with prospective correction in prefrontal cortex in 11 subjects.

RESULTS:

Without prospective motion and shim correction, motion resulted in considerable degradation of MR spectra (broader linewidth, lower signal-to-noise ratio, degraded water suppression). With prospective motion and shim correction, spectral quality remained excellent despite motion. Prospective motion correction alone was not sufficient to prevent degradation of spectral quality.

CONCLUSION:

Prospective motion and B0 shim correction is feasible at 7T and should help improve the robustness of MRS, particularly in motion-prone populations.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Processamento de Imagem Assistida por Computador / Encéfalo / Espectroscopia de Ressonância Magnética / Movimento (Física) Limite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Magn Reson Med Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2019 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Processamento de Imagem Assistida por Computador / Encéfalo / Espectroscopia de Ressonância Magnética / Movimento (Física) Limite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Magn Reson Med Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2019 Tipo de documento: Article