The value of iterative metal artifact reduction algorithms during antenna positioning for CT-guided microwave ablation.
Do, Thuy Duong; Melzig, Claudius; Vollherbst, Dominik F; Pereira, Philippe L; Kauczor, Hans-Ulrich; Kachelrieß, Marc; Sommer, Christof M.
Afiliação
  • Do TD; Clinic for Diagnostic and Interventional Radiology, University Hospital Heidelberg, Heidelberg, Germany.
  • Melzig C; Clinic for Diagnostic and Interventional Radiology, University Hospital Heidelberg, Heidelberg, Germany.
  • Vollherbst DF; Department of Neuroradiology, University Hospital Heidelberg, Heidelberg, Germany.
  • Pereira PL; Clinic for Radiology, Minimally-Invasive Therapies and Nuclear Medicine, SLK Kliniken Heilbronn GmbH, Heilbronn, Germany.
  • Kauczor HU; Clinic for Diagnostic and Interventional Radiology, University Hospital Heidelberg, Heidelberg, Germany.
  • Kachelrieß M; Medical Physics in Radiology, German Cancer Research Center (Dkfz), Heidelberg, Germany.
  • Sommer CM; Clinic for Diagnostic and Interventional Radiology, University Hospital Heidelberg, Heidelberg, Germany.
Int J Hyperthermia ; 36(1): 1223-1232, 2019.
Article em En | MEDLINE | ID: mdl-31814464
ABSTRACT

Objectives:

To compare image quality between filtered back projection (FBP) and iterative reconstruction algorithm and dedicated metal artifact reduction (iMAR) algorithms during antenna positioning for computed tomography-guided microwave ablation (MWA).Materials and

methods:

An MWA antenna was positioned in the liver of five pigs under CT guidance. Different exposure settings (120kVp/200mAs-120kVp/50mAs) and image reconstruction techniques (FBP, iterative reconstruction with and without iMAR) were applied. Quantitative image analysis included density measurements in six positions (e.g., liver in extension of the antenna [ANTENNA] and liver >3 cm away from the antenna [LIVER-1]). Qualitative image analysis included assessment of overall quality, image noise, artifacts at the antenna tip, artifacts in liver parenchyma bordering antenna tip and newly generated artifacts. Two independent observers performed the analyses twice and interreader agreement was compared with Bland-Altman analysis.

Results:

For all exposure and reconstruction settings, density measurements for ANTENNA were significantly higher for the I30-1 iMAR compared with FBP and I30-1 (e.g., 8.3-17.2HU vs. -104.5 to 155.1HU; p ≤ 0.01, respectively). In contrast, for all exposure settings, density measurements for LIVER-1 were comparable between FBP and I30-1 iMAR (e.g., 49.4-50.4HU vs. 50.1-52.5U, respectively). For all exposure and reconstruction settings, subjective image quality for LIVER-1 was better for the I30-1 iMAR algorithm compared with FBP and I30-1. Bland-Altman interobserver agreement was from -0.2 to 0.2 for FBP and iMAR, and Cohen's kappa was 0.74.

Conclusion:

Iterative algorithms I30-1 with iMAR algorithm improves image quality during antenna positioning and placement for CT-guided MWA and is applicable over a range of exposure settings.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Tomografia Computadorizada por Raios X / Artefatos / Técnicas de Ablação / Metais Tipo de estudo: Qualitative_research Limite: Animals / Female / Humans / Male Idioma: En Revista: Int J Hyperthermia Assunto da revista: NEOPLASIAS / TERAPEUTICA Ano de publicação: 2019 Tipo de documento: Article País de afiliação: Alemanha

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Tomografia Computadorizada por Raios X / Artefatos / Técnicas de Ablação / Metais Tipo de estudo: Qualitative_research Limite: Animals / Female / Humans / Male Idioma: En Revista: Int J Hyperthermia Assunto da revista: NEOPLASIAS / TERAPEUTICA Ano de publicação: 2019 Tipo de documento: Article País de afiliação: Alemanha