Biomechanical comparison of knotless vs. knotted suture anchors in the acetabular rim with respect to bone density.
Benca, Emir; Dlaska, Constantin E; DiFranco, Matthew D; Hobusch, Gerhard M; Windhager, Reinhard; Angeli, Thomas; von Skrbensky, Gobert.
Afiliação
  • Benca E; Department of Orthopedics and Trauma Surgery, Medical University of Vienna, Währinger Gürtel 18-20, 1090 Vienna, Austria. Electronic address: emir.benca@muv.ac.at.
  • Dlaska CE; Department of Orthopedics and Trauma Surgery, Medical University of Vienna, Währinger Gürtel 18-20, 1090 Vienna, Austria; Center for Anatomy and Cell Biology, Medical University of Vienna, Währinger Straße 13, 1090 Vienna, Austria.
  • DiFranco MD; Center for Medical Physics and Biomedical Engineering, Medical University of Vienna, Währinger Gürtel 18-20, 1090 Vienna, Austria.
  • Hobusch GM; Department of Orthopedics and Trauma Surgery, Medical University of Vienna, Währinger Gürtel 18-20, 1090 Vienna, Austria.
  • Windhager R; Department of Orthopedics and Trauma Surgery, Medical University of Vienna, Währinger Gürtel 18-20, 1090 Vienna, Austria.
  • Angeli T; Institute of Engineering Design and Product Development, TU Wien, Getreidemarkt 9/307, 1060 Vienna, Austria.
  • von Skrbensky G; Department of Orthopedics and Trauma Surgery, Medical University of Vienna, Währinger Gürtel 18-20, 1090 Vienna, Austria.
Clin Biomech (Bristol, Avon) ; 74: 21-26, 2020 04.
Article em En | MEDLINE | ID: mdl-32109719
ABSTRACT

BACKGROUND:

Acetabular labral tears are managed with suture anchors providing good clinical outcomes. Knotless anchors are easier to use and have a quicker insertion time compared to knotted anchors. The purpose of this study was to compare the biomechanical behavior of two different anchor designs (knotted vs. knotless) in ultimate load testing in correlation with bone density in the acetabular rim.

METHODS:

Eighteen knotted Bio-FASTak and seventeen knotless PushLock anchors (both Arthrex Inc., Naples, FL, USA) were inserted in the rims of two human acetabula, with known bone density distribution. The anchors were subjected to load-to-failure tests. Anchors were additionally tested in solid polyurethane foam with defined densities.

FINDINGS:

The Bio-FASTak group showed higher survival rates at 1, 2, and 3 mm displacement and was able to withstand significantly higher loads at 3 mm displacement (p = 0.031). There was no statistically significant difference in stiffness (p = 0.087), yield- (p = 0.190), and ultimate load (p = 0.222) between the two groups. Only the PushLock group showed correlation between bone volume over total volume (BV/TV) and stiffness (R = 0.750, p = 0.086) and between BV/TV and yield load (R = 0.838, p = 0.037). Experiments on solid polyurethane foam confirmed the correlation between the mechanical properties and tissue density for the same anchor.

INTERPRETATION:

PushLock shows similar biomechanical properties to the Bio-FASTak, but eliminates knot tying and potentially abrasive knots. In addition, biomechanical properties of the PushLock are governed by local bone density.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Densidade Óssea / Âncoras de Sutura / Acetábulo Limite: Humans / Male Idioma: En Revista: Clin Biomech (Bristol, Avon) Assunto da revista: ENGENHARIA BIOMEDICA / FISIOLOGIA Ano de publicação: 2020 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Densidade Óssea / Âncoras de Sutura / Acetábulo Limite: Humans / Male Idioma: En Revista: Clin Biomech (Bristol, Avon) Assunto da revista: ENGENHARIA BIOMEDICA / FISIOLOGIA Ano de publicação: 2020 Tipo de documento: Article