Your browser doesn't support javascript.
loading
Evaluation of Left Main Coronary Artery Using Optical Frequency Domain Imaging and Its Pitfalls.
Roule, Vincent; Rebouh, Idir; Lemaitre, Adrien; Bignon, Mathieu; Ardouin, Pierre; Sabatier, Rémi; Labombarda, Fabien; Blanchart, Katrien; Beygui, Farzin.
Afiliação
  • Roule V; CHU de Caen Normandie, Service de Cardiologie, Caen 14000, France.
  • Rebouh I; Normandie University, UNICAEN, EA 4650 Signalisation, Électrophysiologie et imagerie des lésions d'ischémie-reperfusion myocardique, Caen 14000, France.
  • Lemaitre A; CHU de Caen Normandie, Service de Cardiologie, Caen 14000, France.
  • Bignon M; CHU de Caen Normandie, Service de Cardiologie, Caen 14000, France.
  • Ardouin P; CHU de Caen Normandie, Service de Cardiologie, Caen 14000, France.
  • Sabatier R; CHU de Caen Normandie, Service de Cardiologie, Caen 14000, France.
  • Labombarda F; CHU de Caen Normandie, Service de Cardiologie, Caen 14000, France.
  • Blanchart K; CHU de Caen Normandie, Service de Cardiologie, Caen 14000, France.
  • Beygui F; Normandie University, UNICAEN, EA 4650 Signalisation, Électrophysiologie et imagerie des lésions d'ischémie-reperfusion myocardique, Caen 14000, France.
J Interv Cardiol ; 2020: 4817239, 2020.
Article em En | MEDLINE | ID: mdl-32581660
ABSTRACT

OBJECTIVES:

We aimed to assess the quality of optical frequency domain imaging (OFDI) of the left main (LM) arterial wall and describe and analyse potential artefacts in this setting.

BACKGROUND:

OFDI is increasingly used to assess ambiguous lesions and optimize LM percutaneous coronary intervention. However, its ability to provide artefact-free high-quality images of coronary ostia and large segments such as the LM remains uncertain.

METHODS:

We included 42 consecutive patients who underwent OFDI, including LM imaging. Each OFDI frame was subdivided into four quadrants and analysed. The number of quadrants with artifacts was calculated within the proximal, mid, and distal LM and the first 5 mm of the left anterior descending artery (LAD) and/or left circumflex artery (LCX).

RESULTS:

The quadrants analysis showed an overall artifact rate of 8.9%, mostly out-of-field (45.1%) or residual blood (44.7%) artefacts. Most artifacts were located in the proximal LM (18.6%) with a stepwise reduction of artifact rates towards distal segments (mid LM 5.8%; distal LM 3.6%, ostial LAD 2.6%, and ostial LCX 0%; p < 0.001). While 20 (48.8%) patients had angiographically visible plaques, OFDI showed plaques in 32 patients (76.2%; p=0.007).

CONCLUSION:

OFDI can accurately evaluate the LM and detect and assess angiographically unvisualized atherosclerotic plaques providing accurate assessment of >90% of the quadrants of the LM and the ostia of its bifurcation branches. However, artifacts mainly located in the proximal LM and decreasing distally in a stepwise fashion should be considered in the interpretation of OFDI in this setting.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Doença da Artéria Coronariana / Vasos Coronários / Tomografia de Coerência Óptica / Placa Aterosclerótica Tipo de estudo: Prognostic_studies Limite: Female / Humans / Male / Middle aged Idioma: En Revista: J Interv Cardiol Assunto da revista: CARDIOLOGIA Ano de publicação: 2020 Tipo de documento: Article País de afiliação: França

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Doença da Artéria Coronariana / Vasos Coronários / Tomografia de Coerência Óptica / Placa Aterosclerótica Tipo de estudo: Prognostic_studies Limite: Female / Humans / Male / Middle aged Idioma: En Revista: J Interv Cardiol Assunto da revista: CARDIOLOGIA Ano de publicação: 2020 Tipo de documento: Article País de afiliação: França