Immunoaffinity microflow liquid chromatography/tandem mass spectrometry for the quantitation of PD1 and PD-L1 in human tumor tissues.
Rapid Commun Mass Spectrom
; 34(20): e8896, 2020 Oct 30.
Article
em En
| MEDLINE
| ID: mdl-32666620
ABSTRACT
RATIONALE High tumor expression of programmed cell death protein (PD1) and programmed death-ligand 1 (PD-L1) is thought to be associated with positive clinical outcomes after treatment with anti-PD1 or anti-PD-L1 agents. Several sensitive methods based on immunohistochemistry, ligand binding assay (LBA), and liquid chromatography/mass spectrometry involving the measurement of PD1 and PD-L1 expression have been reported. Here, we expand on the characterization of different tumor types using a highly specific, sensitive, and robust immunoaffinity liquid chromatography/tandem mass spectrometry (IA-LC/MS/MS)-based method for the simultaneous quantitation of PD1 and PD-L1 in tumor tissues. METHODS:
Human tumor tissue samples were homogenized using a Precellys Evolution homogenizer. The samples were incubated with anti-PD1 and anti-PD-L1 capture polyclonal antibodies, which were bound to magnetic beads. Following enrichment, samples were digested with trypsin. A Waters iKEY HSS T3 1.8 um (150 µm × 100 mm) column with a gradient flow rate of 3 µL/min was used for chromatographic separation, and a Waters TQ-S triple quadrupole mass spectrometer was used for detection. Selected reaction monitoring (SRM) transitions with unit resolution for precursor/product ion masses were optimized for PD1 and PD-L1 surrogate peptides.RESULTS:
The surrogate peptides LAAFPEDR for PD1 and FTVTVPK for PD-L1 yielded the most intense SRM transitions at m/z 459.7 > 516.2 and m/z 396.2 > 543.3, respectively, and thus were selected for the quantitation of PD1 and PD-L1. The lower limit of quantitation for PD1 and PD-L1 was 0.062 ng/mL with an assay range up to 10 ng/mL. Using this method, human PD1 and PD-L1 were detected and quantified from four different types of tumor tissues. The data show that PD1 expression level was highly correlated with that of PD-L1 in all tumor tissues analyzed here.CONCLUSIONS:
A highly specific and sensitive immunoaffinity microflow LC/MS/MS method for the simultaneous quantification of PD1 and PD-L1 in tumor tissues was developed and implemented. This method combines the advantage of immuno-capture for analyte enrichment with the high specificity of detection of multiple surrogate peptides by LC/MS/MS. The quantification of PD1 and PD-L1 co-expression in tumor could help evaluate their role in assessing tumor type selection and patient stratification.
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Cromatografia Líquida
/
Espectrometria de Massas em Tandem
/
Antígeno B7-H1
/
Receptor de Morte Celular Programada 1
/
Neoplasias
Limite:
Humans
Idioma:
En
Revista:
Rapid Commun Mass Spectrom
Ano de publicação:
2020
Tipo de documento:
Article
País de afiliação:
Estados Unidos