Your browser doesn't support javascript.
loading
Large-volume electric field-assisted multiphase extraction of malachite green from water samples: A multisample device and method validation.
Viana, Jaime Dos Santos; Caneschi de Freitas, Marina; Botelho, Bruno Gonçalves; Orlando, Ricardo M.
Afiliação
  • Viana JDS; Laboratório de Microfluídica e Separações, LaMS, Departamento de Química, Universidade Federal de Minas Gerais, Belo Horizonte, 30123-970, Minas Gerais, Brazil.
  • Caneschi de Freitas M; Laboratório de Microfluídica e Separações, LaMS, Departamento de Química, Universidade Federal de Minas Gerais, Belo Horizonte, 30123-970, Minas Gerais, Brazil.
  • Botelho BG; Laboratório de Microfluídica e Separações, LaMS, Departamento de Química, Universidade Federal de Minas Gerais, Belo Horizonte, 30123-970, Minas Gerais, Brazil.
  • Orlando RM; Laboratório de Microfluídica e Separações, LaMS, Departamento de Química, Universidade Federal de Minas Gerais, Belo Horizonte, 30123-970, Minas Gerais, Brazil. Electronic address: orlandoricardo@hotmail.com.
Talanta ; 222: 121540, 2021 Jan 15.
Article em En | MEDLINE | ID: mdl-33167248
In this work, a new large-volume multiphase, multi-sample electroextraction device was developed and applied to selectively extract malachite green (MG) from water samples. This device was easily constructed with ordinary materials and capable of extracting ten samples simultaneously, obtaining MG preconcentrated on a solid support, to fit into a pipette tip. A multi-well plate was applied to extract MG from aquaculture water samples, and the extracts containing the desorbed MG were analysed by LC-DAD and LC-MS/MS. The signals from both detectors were used in two independent validation procedures. Linearity, matrix effect, selectivity, precision, trueness, and limits of detection and quantification were all evaluated. For both detectors, linearity was demonstrated in the range of 0.5-5 µg L-1 (R2 > 0.98). Matrix effect was insignificant for LC-DAD only, and the average preconcentration factor was about 60 times. Recoveries ranged from 94 to 113% for LC-DAD and 95-115% for LC-MS/MS analysis. ANOVA was applied to estimate the standard deviation under repeatability (6.96-8.61% for LC-DAD and 5.98-7.41% for LC-MS/MS) and within-reproducibility (6.96-8.61% for LC-DAD and 6.56-7.41% for LC-MS/MS) conditions. The limits of detection and quantification for LC-MS/MS analysis were 4.29 and 28.74 ng L-1, respectively, while, for LC-DAD, these limits were 14.29 and 95.81 ng L-1, respectively. The results demonstrated that the developed method was suitable for determining MG in water samples, and the large-volume multiphase, multi-sample electroextraction device proved to be a powerful sample preparation technique to obtain high clean-up and large preconcentration levels, which are of paramount importance for environmental applications.
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Revista: Talanta Ano de publicação: 2021 Tipo de documento: Article País de afiliação: Brasil

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Revista: Talanta Ano de publicação: 2021 Tipo de documento: Article País de afiliação: Brasil