Your browser doesn't support javascript.
loading
Coagulation complications after conversion from roller to centrifugal pump in neonatal and pediatric extracorporeal membrane oxygenation.
Erdem, Özge; Kuiper, Jan Willem; Houmes, Robert Jan; van Ommen, Cornelia Heleen; van Rosmalen, Joost; Tibboel, Dick; Wildschut, Enno Diederik.
Afiliação
  • Erdem Ö; Intensive Care and Department of Pediatric Surgery, Erasmus University Medical Center - Sophia Children's Hospital, Dr. Molewaterplein 40, 3015GD, Rotterdam, the Netherlands.
  • Kuiper JW; Intensive Care and Department of Pediatric Surgery, Erasmus University Medical Center - Sophia Children's Hospital, Dr. Molewaterplein 40, 3015GD, Rotterdam, the Netherlands. Electronic address: j.kuiper@erasmusmc.nl.
  • Houmes RJ; Intensive Care and Department of Pediatric Surgery, Erasmus University Medical Center - Sophia Children's Hospital, Dr. Molewaterplein 40, 3015GD, Rotterdam, the Netherlands.
  • van Ommen CH; Department of Pediatric Hematology, Erasmus University Medical Center - Sophia Children's Hospital, Rotterdam, the Netherlands.
  • van Rosmalen J; Department of Biostatistics, Erasmus University Medical Center, Rotterdam, the Netherlands.
  • Tibboel D; Intensive Care and Department of Pediatric Surgery, Erasmus University Medical Center - Sophia Children's Hospital, Dr. Molewaterplein 40, 3015GD, Rotterdam, the Netherlands.
  • Wildschut ED; Intensive Care and Department of Pediatric Surgery, Erasmus University Medical Center - Sophia Children's Hospital, Dr. Molewaterplein 40, 3015GD, Rotterdam, the Netherlands.
J Pediatr Surg ; 56(8): 1378-1385, 2021 Aug.
Article em En | MEDLINE | ID: mdl-33279215
ABSTRACT
BACKGROUND/

PURPOSE:

Coagulation complications are frequent, unwanted occurrences in extracorporeal membrane oxygenation (ECMO) treatment, possibly influenced by the pump in the ECMO-circuit. We hypothesized that fewer complications would occur with a smaller, heparin-coated ECMO system with a centrifugal pump (CP) than with one with a roller pump (RP) and that after conversion, complication rates would decrease over time.

METHODS:

This single-center, retrospective chart study included all first neonatal and pediatric ECMO runs between 2009 and 2015. Differences between groups were assessed with Mann-Whitney U tests and Kruskal-Wallis tests. Determinants of complication rates were evaluated through Poisson regression models. The CP group was divided into three consecutive groups to assess whether complication rates decreased over time.

RESULTS:

The RP group comprised 90 ECMO runs and the CP group 82. Hemorrhagic complication rates were significantly higher with the CP than with the RP, without serious therapeutic consequences, while thrombotic complications rates were unaffected. Intracranial hemorrhage rates and coagulation-related mortality rates were similar. Gained experience with the CP did not improve complication rates or survival over time.

CONCLUSIONS:

Although the CP seems safe, it does not seem beneficial over the RP. Further research is warranted on how pump type affects coagulation, taking into account the severity and implications of coagulation complications. LEVEL OF EVIDENCE Level III.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Trombose / Oxigenação por Membrana Extracorpórea Tipo de estudo: Etiology_studies / Observational_studies / Risk_factors_studies Limite: Child / Humans / Newborn Idioma: En Revista: J Pediatr Surg Ano de publicação: 2021 Tipo de documento: Article País de afiliação: Holanda

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Trombose / Oxigenação por Membrana Extracorpórea Tipo de estudo: Etiology_studies / Observational_studies / Risk_factors_studies Limite: Child / Humans / Newborn Idioma: En Revista: J Pediatr Surg Ano de publicação: 2021 Tipo de documento: Article País de afiliação: Holanda