Your browser doesn't support javascript.
loading
Flexible Versus Rigid Laryngoscopy: A Prospective, Blinded Comparison of Image Quality.
Boles, Roger W; Gao, William Z; Johns, Michael M; Daniero, James J; Grant, Nazaneen N; Rubin, Adam D; Bhatt, Neel K; Hapner, Edie R; O'Dell, Karla.
Afiliação
  • Boles RW; Keck School of Medicine of USC, Los Angeles, California.
  • Gao WZ; Department of Otolaryngology-Head and Neck Surgery, Georgetown University Medical Center, Washington, District of Columbia.
  • Johns MM; Department of Otolaryngology-Head and Neck Surgery, University of Southern California, Los Angeles, California.
  • Daniero JJ; Department of Otolaryngology-Head and Neck Surgery, University of Virginia Health System, Charlottesville, Virginia.
  • Grant NN; Department of Otolaryngology-Head and Neck Surgery, Georgetown University Medical School, Washington, District of Columbia.
  • Rubin AD; Lakeshore Ears, Nose, Throat Center, St. Clair Shores, Michighan.
  • Bhatt NK; Department of Otolaryngology-Head and Neck Surgery, University of Southern California, Los Angeles, California.
  • Hapner ER; Department of Otolaryngology-Head and Neck Surgery, University of Southern California, Los Angeles, California.
  • O'Dell K; Department of Otolaryngology-Head and Neck Surgery, University of Southern California, Los Angeles, California. Electronic address: Karla.ODell@med.usc.edu.
J Voice ; 37(3): 440-443, 2023 May.
Article em En | MEDLINE | ID: mdl-33775470
ABSTRACT

OBJECTIVE:

To compare flexible distal-chip laryngoscopy (FDL) and rigid telescopic laryngoscopy (RTL) in image quality and diagnostic ability. STUDY

DESIGN:

Prospective cohort study; blinded comparison.

METHODS:

Eighteen normal adult subjects were recruited to undergo both FDL and RTL and normalized videos were recorded. Three blinded laryngologists compared the videos for color fidelity, illumination, resolution, and vascularity, and indicated superiority with FDL, RTL, or no difference. Raters also reported if an abnormality was seen and in which video it was better visualized. Videos for two subjects were repeated to assess intra-rater reliability, making 20 video comparisons across 3 raters for a total of 60 ratings. Differences in responses were analyzed via Mann-Whitney U and Pearson Χ2. Inter-rater reliability was assessed via Fleiss' kappa, and intra-rater reliability was assessed via percent agreement.

RESULTS:

RTL was rated superior in all categories of image quality (47 vs 5 vs 8, P < 0.01; 47 vs 7 vs 6, P < 0.01; 51 vs 5 vs 4, P<0.01; 44 vs 9 vs 7, P < 0.01, respectively). An abnormality was seen 33 times with both modalities and 6 times with RTL only. When seen with both modalities, visualization was superior in RTL compared with FDL (29 vs 4, P <0.01).

CONCLUSIONS:

There was significant superiority of RTL in all categories of image quality, with slight inter-rater agreement for color fidelity, resolution, and vascularity. RTL was also significantly better for visualization of abnormalities. These findings suggest superior image quality in RTL compared with FDL, but further research is required to determine if this difference is clinically significant.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Iluminação / Laringoscopia Tipo de estudo: Observational_studies Limite: Adult / Humans Idioma: En Revista: J Voice Assunto da revista: OTORRINOLARINGOLOGIA Ano de publicação: 2023 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Iluminação / Laringoscopia Tipo de estudo: Observational_studies Limite: Adult / Humans Idioma: En Revista: J Voice Assunto da revista: OTORRINOLARINGOLOGIA Ano de publicação: 2023 Tipo de documento: Article