Your browser doesn't support javascript.
loading
Evaluation and Comparison with Other High-Sensitivity Methods of Analytical Performance and Measured Values of a New Laboratory Test for Cardiac Troponin I Assay.
Masotti, Silvia; Musetti, Veronica; Prontera, Concetta; Storti, Simona; Passino, Claudio; Clerico, Aldo.
Afiliação
  • Masotti S; Scuola Superiore Sant'Anna and Fondazione CNR Regione Toscana G. Monasterio, Pisa, Italy.
  • Musetti V; Scuola Superiore Sant'Anna and Fondazione CNR Regione Toscana G. Monasterio, Pisa, Italy.
  • Prontera C; Scuola Superiore Sant'Anna and Fondazione CNR Regione Toscana G. Monasterio, Pisa, Italy.
  • Storti S; Scuola Superiore Sant'Anna and Fondazione CNR Regione Toscana G. Monasterio, Pisa, Italy.
  • Passino C; Scuola Superiore Sant'Anna and Fondazione CNR Regione Toscana G. Monasterio, Pisa, Italy.
  • Clerico A; Scuola Superiore Sant'Anna and Fondazione CNR Regione Toscana G. Monasterio, Pisa, Italy.
J Appl Lab Med ; 6(5): 1237-1250, 2021 09 01.
Article em En | MEDLINE | ID: mdl-33829255
ABSTRACT
BAKGROUND The aim of this study was to evaluate both analytical characteristics and clinical results of a new chemiluminescent method for the measurement of cardiac troponin I (cTnI), named VITROS ® High Sensitivity Troponin I Assay, using the VITROS® 3600 automated platform. The results found with this new method were compared to those observed with hs-cTnI ARCHITECT and ECLIA hs-cTnT ELECSYS methods.

METHODS:

For evaluation of analytical performance and comparison of clinical results, plasma samples (lithium-heparin), were collected from apparently healthy subjects and patients with cardiovascular diseases.

RESULTS:

The hs-cTnI VITROS method showed values for limit of blank (LoB 0.33 ng/L), limit of detection (LoD, 0.91 ng/L), limit of quantifications at 20% (LoQ 20% CV, 1.82 ng/L), and 10% (LoQ 10% CV, 4,74 ng/L), which are comparable to those previously reported for other hs-cTnI methods. Moreover, the clinical results of the hs-cTnI VITROS method were found to be closely correlated to those of hs-cTnI ARCHITECT (R = 0,9883, N = 198) and ECLIA hs-cTnT Elecsys (R = 0,9704, N = 293) methods.

CONCLUSIONS:

The hs-cTnI VITROS method shows analytical performance comparable to other cTnI and cTnT assay. The results of this study confirm that there are significant systematic differences among hs-cTnI methods. Further multicenter studies using larger reference populations are needed in order to obtain a better estimation, especially of the 99° percentile URL values categorized for sex and age of hs-cTnI and hs-cTnT methods.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Troponina I Tipo de estudo: Clinical_trials / Diagnostic_studies Limite: Humans Idioma: En Revista: J Appl Lab Med Ano de publicação: 2021 Tipo de documento: Article País de afiliação: Itália

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Troponina I Tipo de estudo: Clinical_trials / Diagnostic_studies Limite: Humans Idioma: En Revista: J Appl Lab Med Ano de publicação: 2021 Tipo de documento: Article País de afiliação: Itália