Radiofrequency-induced heating of broken and abandoned implant leads during magnetic resonance examinations.
Yao, Aiping; Goren, Tolga; Samaras, Theodoros; Kuster, Niels; Kainz, Wolfgang.
Afiliação
  • Yao A; Foundation of Research on Information Technologies in Society (IT'IS), Zurich, Switzerland.
  • Goren T; Foundation of Research on Information Technologies in Society (IT'IS), Zurich, Switzerland.
  • Samaras T; Department of Physics, Aristotle University of Thessaloniki, Thessaloniki, Greece.
  • Kuster N; Foundation of Research on Information Technologies in Society (IT'IS), Zurich, Switzerland.
  • Kainz W; Department of Information Technology and Electrical Engineering, ETH Zurich, Zurich, Switzerland.
Magn Reson Med ; 86(4): 2156-2164, 2021 10.
Article em En | MEDLINE | ID: mdl-34080721
ABSTRACT

PURPOSE:

The risks of RF-induced heating of active implantable medical device (AIMD) leads during MR examinations must be well understood and realistically assessed. In this study, we evaluate the potential additional risks of broken and abandoned (cut) leads.

METHODS:

First, we defined a generic AIMD with a metallic implantable pulse generator (IPG) and a 100-cm long lead containing 1 or 2 wires. Next, we numerically estimated the deposited in vitro lead-tip power for an intact lead, as well as with wire breaks placed at 10 cm intervals. We studied the effect of the break size (wire gap width), as well as the presence of an intact wire parallel to the broken wire, and experimentally validated the numeric results for the configurations with maximum deposited in vitro lead-tip power. Finally, we performed a Tier 3 assessment of the deposited in vivo lead-tip power for the intact and broken lead in 4 high resolution virtual population anatomic models for over 54,000 MR examination scenarios.

RESULTS:

The enhancement of the deposited lead-tip power for the broken leads, compared to the intact lead, reached 30-fold in isoelectric exposure, and 16-fold in realistic clinical exposures. The presence of a nearby intact wire, or even a nearby broken wire, reduced this enhancement factor to <7-fold over the intact lead.

CONCLUSION:

Broken and abandoned leads can pose increased risk of RF-induced lead-tip heating to patients undergoing MR examinations. The potential enhancement of deposited in vivo lead-tip power depends on location and type of the wire break, lead design, and clinical routing of the lead, and should be carefully considered when performing risk assessment for MR examinations and MR conditional labeling.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Imageamento por Ressonância Magnética / Calefação Tipo de estudo: Prognostic_studies / Risk_factors_studies Limite: Humans Idioma: En Revista: Magn Reson Med Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2021 Tipo de documento: Article País de afiliação: Suíça

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Imageamento por Ressonância Magnética / Calefação Tipo de estudo: Prognostic_studies / Risk_factors_studies Limite: Humans Idioma: En Revista: Magn Reson Med Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2021 Tipo de documento: Article País de afiliação: Suíça