Your browser doesn't support javascript.
loading
Whether sutures reduce the graft laceration caused by interference screw in anterior cruciate ligament reconstruction? A biomechanical study in vitro.
Teng, Yuanjun; Zhang, Xiaohui; Da, Lijun; Hu, Jie; Wang, Hong; Han, Hua; Wu, Meng; Zhang, Shifeng; Xia, Yayi.
Afiliação
  • Teng Y; Department of Orthopaedics, Lanzhou University Second Hospital, Lanzhou University, No. 82 Cuiyingmen, Chengguan District, 730030, Lanzhou City, Gansu Province, People's Republic of China.
  • Zhang X; Orthopaedics Key Laboratory of Gansu Province, Lanzhou University Second Hospital, Lanzhou University, 730030, Lanzhou City, Gansu Province, People's Republic of China.
  • Da L; Department of Orthopaedics, Lanzhou University Second Hospital, Lanzhou University, No. 82 Cuiyingmen, Chengguan District, 730030, Lanzhou City, Gansu Province, People's Republic of China.
  • Hu J; Department of Oncology, Lanzhou University Second Hospital, Lanzhou University, 730030, Lanzhou City, Gansu Province, People's Republic of China.
  • Wang H; The Second Clinical Medical College, Lanzhou University, 730030, Lanzhou City, Gansu Province, People's Republic of China.
  • Han H; Department of Orthopaedics, Lanzhou University Second Hospital, Lanzhou University, No. 82 Cuiyingmen, Chengguan District, 730030, Lanzhou City, Gansu Province, People's Republic of China.
  • Wu M; Orthopaedics Key Laboratory of Gansu Province, Lanzhou University Second Hospital, Lanzhou University, 730030, Lanzhou City, Gansu Province, People's Republic of China.
  • Zhang S; Department of Orthopaedics, Lanzhou University Second Hospital, Lanzhou University, No. 82 Cuiyingmen, Chengguan District, 730030, Lanzhou City, Gansu Province, People's Republic of China.
  • Xia Y; Department of Orthopaedics, Lanzhou University Second Hospital, Lanzhou University, No. 82 Cuiyingmen, Chengguan District, 730030, Lanzhou City, Gansu Province, People's Republic of China.
BMC Musculoskelet Disord ; 22(1): 571, 2021 Jun 22.
Article em En | MEDLINE | ID: mdl-34158018
ABSTRACT

BACKGROUND:

Interference screw is commonly used for graft fixation in anterior cruciate ligament (ACL) reconstruction. However, previous studies had reported that the insertion of interference screws significantly caused graft laceration. The purposes of this study were to (1) quantitatively evaluate the graft laceration from one single insertion of PEEK interference screws; and (2) determine whether different types of sutures reduced the graft laceration after one single insertion of interference screws in ACL reconstruction.

METHODS:

The in-vitro ACL reconstruction model was created using porcine tibias and bovine extensor digitorum tendons of bovine hind limbs. The ends of grafts were sutured using three different sutures, including the bioabsorbable, Ethibond and ultra-high molecular weight polyethylene (UHMWPE) sutures. Poly-ether-ether-ketone (PEEK) interference screws were used for tibial fixation. This study was divided into five groups (n = 10 in each group) the non-fixed group, the non-sutured group, the absorbable suture group, the Ethibond suture group and the UHMWPE suture group. Biomechanical tests were performed using the mode of pull-to-failure loading tests at 10 mm/min. Tensile stiffness (newtons per millimeter), energy absorbed to failure (in joules) and ultimate load (newtons) were recorded for analysis.

RESULTS:

All prepared tendons and bone specimens showed similar characteristics (length, weight, and pre-tension of the tendons, tibial bone mineral density) among all groups (P > 0.05). The biomechanical tests demonstrated that PEEK interference screws significantly caused the graft laceration (P < 0.05). However, all sutures (the bioabsorbable, Ethibond and UHMWPE sutures) did not reduce the graft laceration in ACL reconstruction (P > 0.05).

CONCLUSIONS:

Our biomechanical study suggested that the ultimate failure load of grafts was reduced of approximately 25 % after one single insertion of a PEEK interference screw in ACL reconstruction. Suturing the ends of the grafts using different sutures (absorbable, Ethibond and UHMWPE sutures) did not decrease the graft laceration caused by interference screws.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Lacerações / Reconstrução do Ligamento Cruzado Anterior Tipo de estudo: Etiology_studies / Prognostic_studies Limite: Animals Idioma: En Revista: BMC Musculoskelet Disord Assunto da revista: FISIOLOGIA / ORTOPEDIA Ano de publicação: 2021 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Lacerações / Reconstrução do Ligamento Cruzado Anterior Tipo de estudo: Etiology_studies / Prognostic_studies Limite: Animals Idioma: En Revista: BMC Musculoskelet Disord Assunto da revista: FISIOLOGIA / ORTOPEDIA Ano de publicação: 2021 Tipo de documento: Article